CLIOwin 6.5 PCI User's Manual - Audiomatica Srl
CLIOwin 6.5 PCI User's Manual - Audiomatica Srl
CLIOwin 6.5 PCI User's Manual - Audiomatica Srl
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
11.3 A BRIEF DESCRIPTION ON SETTINGS EFFECTS<br />
11.3.1 STEPPED VS. NOT STEPPED<br />
Although Measuring speed increases, use of a not stepped sweep can adversely affect<br />
measuring results in several circumstances. As an example that should make clear also<br />
other situation, let’s see what happens measuring the impedance of a woofer in Internal<br />
or Constant Current Mode. Please refer to Measuring Impedance for more information<br />
on this topic. In both conditions the loudspeaker is driven from a high impedance source<br />
and its damping is only mechanical. Fig.11.1 shows a 6" woofer, driven by a 100Ohm<br />
output impedance generator delivering a sinusoidal burst 200ms long, at its resonance<br />
frequency. When the excitation stops, the device continues to move, and therefore<br />
produce back electromotive force voltage for more than 50 ms. Something very similar<br />
happens at start-up. In this situation, if CLIO is set in not stepped mode, it will acquire<br />
the previously played frequency together with the actual.<br />
0.200<br />
CLIO<br />
V<br />
0.120<br />
0.040<br />
-0.040<br />
-0.120<br />
-0.200<br />
0.00 50 100 150 200 250 300 350 400 ms 450 500<br />
Figure 11.1<br />
Fig.11.2 shows the same driver impedance taken both in stepped (black) and not<br />
stepped mode (red). The not stepped curve is simply wrong. Conclusion: use always<br />
stepped mode if the behaviour of the device to be measured is unknown.<br />
25.0<br />
CLIO<br />
180.0<br />
Ohm<br />
Deg<br />
20.0 108.0<br />
15.0 36.0<br />
10.0 -36.0<br />
5.0 -108.0<br />
0.0 -180.0<br />
10 100 1k Hz<br />
10k 20k<br />
Figure 11.2<br />
110 Chapter 11 - Sinusoidal