11. Hyperhydrosis palmoplantaris<br />

Objective : normalization of<br />

sympathicus<br />

Patient position : sitting, with feet/h<strong>and</strong>s in<br />

two-cell bath<br />

Electrode positions : electrodes are placed at<br />

the bottom of the bath<br />

compartments, sponges<br />

are placed on the<br />

electrodes, after which the<br />

feet/h<strong>and</strong>s are placed on<br />

the sponges<br />

Current type : <strong>medium</strong>-frequency direct<br />

current (reverse polarity<br />

halfway through)<br />

Output characteristic : CC<br />

<strong>Frequency</strong> : -<br />

Phase time : -<br />

Burst : -<br />

<strong>Frequency</strong> modulation : -<br />

Sweep mode : -<br />

Amplitude : sensory level<br />

Treatment time : 16 minutes total<br />

<strong>Frequency</strong> of treatment : three times weekly<br />

Note : sponges must overlap<br />

treated area by a least 1.5<br />

cm<br />

12. Lateral epicondylitis<br />

Objective : introduction of histamin by<br />

iontophoresis<br />

Patient position : sitting<br />

Electrode positions : anode at the level of the<br />

origin of the extensor carpi<br />

radialis (longus or brevis)<br />

cathode (indifferent)<br />

elsewhere on body<br />

Current type : <strong>medium</strong>-frequency direct<br />

current<br />

Output characteristic : CC<br />

<strong>Frequency</strong> : -<br />

Phase time : -<br />

Burst : -<br />

<strong>Frequency</strong> modulation : -<br />

Sweep mode : -<br />

Amplitude : not more than 10 mA, <strong>and</strong><br />

not more than 0.2 mA/cm²<br />

Treatment time : maximum 5 minutes<br />

<strong>Frequency</strong> of treatment : twice per week<br />

Note : • do not use a sponge<br />

between the anode <strong>and</strong><br />

the skin, but an<br />

absorbent gauze<br />

soaked in histamin<br />

solution (max. 0.1 %)<br />

• the indifferent<br />

electrode should<br />

always be larger than<br />

the active electrode<br />

45