11. Hyperhydrosis palmoplantaris<br />
Objective : normalization of<br />
sympathicus<br />
Patient position : sitting, with feet/h<strong>and</strong>s in<br />
two-cell bath<br />
Electrode positions : electrodes are placed at<br />
the bottom of the bath<br />
compartments, sponges<br />
are placed on the<br />
electrodes, after which the<br />
feet/h<strong>and</strong>s are placed on<br />
the sponges<br />
Current type : <strong>medium</strong>-frequency direct<br />
current (reverse polarity<br />
halfway through)<br />
Output characteristic : CC<br />
<strong>Frequency</strong> : -<br />
Phase time : -<br />
Burst : -<br />
<strong>Frequency</strong> modulation : -<br />
Sweep mode : -<br />
Amplitude : sensory level<br />
Treatment time : 16 minutes total<br />
<strong>Frequency</strong> of treatment : three times weekly<br />
Note : sponges must overlap<br />
treated area by a least 1.5<br />
cm<br />
12. Lateral epicondylitis<br />
Objective : introduction of histamin by<br />
iontophoresis<br />
Patient position : sitting<br />
Electrode positions : anode at the level of the<br />
origin of the extensor carpi<br />
radialis (longus or brevis)<br />
cathode (indifferent)<br />
elsewhere on body<br />
Current type : <strong>medium</strong>-frequency direct<br />
current<br />
Output characteristic : CC<br />
<strong>Frequency</strong> : -<br />
Phase time : -<br />
Burst : -<br />
<strong>Frequency</strong> modulation : -<br />
Sweep mode : -<br />
Amplitude : not more than 10 mA, <strong>and</strong><br />
not more than 0.2 mA/cm²<br />
Treatment time : maximum 5 minutes<br />
<strong>Frequency</strong> of treatment : twice per week<br />
Note : • do not use a sponge<br />
between the anode <strong>and</strong><br />
the skin, but an<br />
absorbent gauze<br />
soaked in histamin<br />
solution (max. 0.1 %)<br />
• the indifferent<br />
electrode should<br />
always be larger than<br />
the active electrode<br />
45