Defibrillation Wave<strong>for</strong>m<br />
category<br />
specifications<br />
Laerdal Medical<br />
wave<strong>for</strong>m parameters<br />
Biphasic truncated exponential. Wave<strong>for</strong>m parameters are automatically<br />
adjusted as a function of patient defibrillation impedance. In the diagram at left,<br />
D is the duration of phase 1 and E is the duration of phase 2 of the wave<strong>for</strong>m, F<br />
is the interphase delay (500 μs), and Ip is the peak current.<br />
The HeartStart delivers shocks to load impedances from 25 to 180 ohms. The<br />
duration of each phase of the wave<strong>for</strong>m is dynamically adjusted based on<br />
delivered charge, in order to compensate <strong>for</strong> patient impedance variations, as<br />
shown below:<br />
adult defibrillation<br />
load phase 1 phase 2 peak delivered<br />
resistance (Ω) duration (ms) duration (ms) current (A) energy (J)<br />
25 2.8 2.8 65 128<br />
50 4.5 4.5 40 150<br />
75 6.25 5.0 30 155<br />
100 8.0 5.3 24 157<br />
125 9.65 6.4 21 159<br />
150 11.5 7.7 18 160<br />
175 12.0 8.0 16 158<br />
pediatric defibrillation<br />
(using M5072A infant/child reduced-energy defibrillator pads)<br />
load phase 1 phase 2 peak delivered<br />
resistance (Ω) duration (ms) duration (ms) current (A) energy (J)<br />
25 4.1 2.8 28 35<br />
50 5.1 3.4 20 46<br />
75 6.2 4.1 15 52<br />
100 7.2 4.8 12 54<br />
125 8.3 5.5 10 56<br />
150 9.0 6.0 9 57<br />
175 9.0 6.0 8 55<br />
E