If the 2-Minute Audio Off mode is selected in the <strong>Monitor</strong> Configuration menu, pressing the<br />
SILENCE/RESET pushbutton will temporarily deactivate the audio alarms for two (2) minutes.<br />
The visual indicator next to the SILENCE/RESET pushbutton will illuminate continuously and the<br />
message “2Min Aud” will be displayed along with the “Ready” message in the Message Window<br />
as a reminder.<br />
NOTE:<br />
If the Message Window is currently displaying a time stamp from a previously taken NIBP<br />
measurement, the “2Min Aud” message will alternately be displayed with the time stamp<br />
message.<br />
When the 2-Minute Audio Off is enabled, the monitor will not alarm for patient related High and<br />
Low alarms, NIBP Application Error and SpO 2 Probe Alarm. The monitor will alarm for Low<br />
Battery and Dead Battery.<br />
To re-activate the audio alarm, press the SILENCE/RESET pushbutton again. The visual<br />
indicators will go out. If the button is not pressed, audible alarms will be re-armed automatically<br />
after two (2) minutes. Refer to Page 101, AUDIO ALARM SILENCE (SILENCE/RESET<br />
Pushbutton) for more information.<br />
PERMANENT AUDIO ALARM SILENCE<br />
NOTE:<br />
Enabling the Permanent Audio Alarm Off mode can only be accomplished after all active alarm<br />
conditions have been addressed. To clear an alarm, refer to Page 95, CLEARING ALARMS for<br />
more information.<br />
If the Permanent Audio Off mode is selected in the <strong>Monitor</strong> Configuration menu, pressing the<br />
SILENCE/RESET pushbutton will deactivate the audio alarms. The visual indicator next to the<br />
SILENCE/RESET pushbutton will flash at a rate of one (1) second “ON” and one (1) second<br />
“OFF” and the message “Perm Aud” will be displayed along with the “Ready” message in the<br />
Message Window as a reminder.<br />
NOTE:<br />
If the Message Window is currently displaying a time stamp from a previously taken NIBP<br />
measurement, the “Perm Aud” message will alternately be displayed with the time stamp<br />
message.<br />
When enabled, the monitor will not alarm for patient related High and Low alarms, NIBP<br />
Application Error and SpO 2 Probe Alarm. The monitor will alarm for Low Battery and Dead<br />
Battery.<br />
To re-activate the audio alarm, press the SILENCE/RESET pushbutton again. The visual<br />
indicators will go out. Refer to Page 101 AUDIO ALARM SILENCE (SILENCE/RESET<br />
Pushbutton) for more information.<br />