<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

CLEARING ALARMS<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> provides to the user an audible and visual indication for both patient<br />

and equipment alarm conditions.<br />

NIBP PATIENT ALARMS<br />

During an active NIBP Patient Alarm Limit Violation (Low/High Systolic or Low/High Diastolic),<br />

the monitor flashes the LED display of the parameter in alarm and provides a Patient Alarm Limit<br />

Violation audible tone.<br />

To acknowledge the alarm, press the SILENCE/RESET pushbutton.<br />

The monitor silences the audible tone and the parameter in alarm stops flashing.<br />

HIGH / LOW %SpO 2 ALARMS<br />

During either a High or Low %SpO 2 Alarm Limit Violation, the monitor flashes the %SpO 2 LED<br />

display, provides a Patient Alarm Limit Violation audible tone and displays the appropriate<br />

message (SpO 2 Hi or SpO 2 Lo) in the Message Window.<br />

To acknowledge the alarm and temporarily mute the audio, press the SILENCE/RESET<br />

pushbutton.<br />

The monitor silences the audible tone for a maximum of fifteen (15) seconds, the parameter in<br />

alarm continues to flash on the front panel and the appropriate message will be displayed in the<br />

Message Window. If the current patient alarm condition becomes inactive during the 15<br />

seconds and then recurs OR a different patient alarm condition occurs during the 15 seconds,<br />

the audio alarm is reactivated. If the alarm condition continues uninterrupted for the 15 seconds,<br />

the audio alarm is reactivated.<br />

SpO 2 PULSE RATE ALARM<br />

During a SpO 2 Pulse Rate Alarm Limit Violation, the monitor flashes the ♥BPM LED display and<br />

provides a Patient Alarm Limit Violation audible tone.<br />

To acknowledge the alarm and temporarily mute the audio, press the SILENCE/RESET<br />

pushbutton.<br />

The monitor silences the audible tone for a maximum of fifteen (15) seconds and the parameter<br />

in alarm continues to flash on the front panel. If the current patient alarm condition becomes<br />

inactive during the 15 seconds and then recurs OR a different patient alarm condition occurs<br />

during the 15 seconds, the audio alarm is reactivated. If the alarm condition continues<br />

uninterrupted for the 15 seconds, the audio alarm is reactivated.<br />