<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

ALARM LIMIT VALUES<br />

There are two sets of alarm limit parameters, one for Adult/Pediatric and one for Neonatal. The<br />

Alarm Limits pushbutton will operate on the parameters for the mode the monitor is currently in<br />

(ADULT / NEO). Table 5, lists the Alarm Limit Default Values used by the <strong>CASMED</strong> <strong>740</strong><br />

<strong>Monitor</strong>.<br />

ADULT Systolic Diastolic MAP Pulse %SpO 2<br />

High 240 130 180 220 OFF<br />

Low OFF OFF OFF OFF 88<br />

NEONATAL Systolic Diastolic MAP Pulse %SpO 2<br />

High 120 80 100 220 96<br />

Low OFF OFF OFF OFF 88<br />

Table 3: Default Alarm Values<br />

NOTE:<br />

Switching modes from Adult to Neonate or Neonate to Adult, shall recall the last stored values<br />

for patient alarm limit.<br />

NOTE:<br />

Patient alarms for Systolic, Diastolic and MAP values are produced at the time the measurement<br />

is taken. Alarms for %SpO 2 have a ten (10) second delay. Alarm for Pulse Rate will be<br />

immediate if taken from the NIBP or have a ten (10) second delay if taken from the SpO 2 signal.<br />

21-02-0298 REV. 00 12/06<br />