<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />
ALARM LIMIT VALUES<br />
There are two sets of alarm limit parameters, one for Adult/Pediatric and one for Neonatal. The<br />
Alarm Limits pushbutton will operate on the parameters for the mode the monitor is currently in<br />
(ADULT / NEO). Table 5, lists the Alarm Limit Default Values used by the <strong>CASMED</strong> <strong>740</strong><br />
<strong>Monitor</strong>.<br />
ADULT Systolic Diastolic MAP Pulse %SpO 2<br />
High 240 130 180 220 OFF<br />
Low OFF OFF OFF OFF 88<br />
NEONATAL Systolic Diastolic MAP Pulse %SpO 2<br />
High 120 80 100 220 96<br />
Low OFF OFF OFF OFF 88<br />
Table 3: Default Alarm Values<br />
NOTE:<br />
Switching modes from Adult to Neonate or Neonate to Adult, shall recall the last stored values<br />
for patient alarm limit.<br />
NOTE:<br />
Patient alarms for Systolic, Diastolic and MAP values are produced at the time the measurement<br />
is taken. Alarms for %SpO 2 have a ten (10) second delay. Alarm for Pulse Rate will be<br />
immediate if taken from the NIBP or have a ten (10) second delay if taken from the SpO 2 signal.<br />
21-02-0298 REV. 00 12/06<br />