CHANGING ALARM LIMITS<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> incorporates a two (2) step process for reviewing and changing<br />

Alarm Limit values.<br />

Press the ALARM LIMITS pushbutton once and the monitor’s front panel displays indicate the<br />

current values of all High Alarm Limit values, for two (2) seconds then toggles to indicate all Low<br />

Alarm Limit values for two (2) seconds. This reviewing of all High and Low Alarm Limit values<br />

continues for thirty (30) seconds if no other pushbutton is pressed.<br />

To change the value of an Alarm Limit, press the ALARM LIMITS pushbutton again. The Alarm<br />

Limit value is displayed in its corresponding display window and the Message Window will<br />

display the limit’s name.<br />

Use the ARROW UP or ARROW DOWN pushbuttons to change the limit value. Use the<br />

ALARM LIMITS pushbutton key to advance onto the next parameter value.<br />

Press the CANCEL pushbutton to return to "Ready" or after thirty (30) seconds of button<br />

inactivity the monitor will automatically return to "Ready". The alarm value(s) set will now be<br />

used until power is turned “OFF”.<br />

SAVING ALARM LIMITS<br />

To save the Alarm Limit values to non-volatile memory, press and hold the ALARM LIMITS<br />

pushbutton for two (2) seconds while in the Alarm Limits set value mode. The Message Window<br />

will display "Save? N".<br />

Press either the ARROW UP or ARROW DOWN pushbuttons will display “Save? Y” in the<br />

Message Window. Press the ALARM LIMIT pushbutton a final time.<br />

The Message Window will display “Saving” and the current value(s) will be saved in memory.<br />

When finished, press the CANCEL pushbutton to return to the “Ready” state. These values will<br />

be retained in memory even after the monitor is turned “OFF”.<br />

RESTORE FACTORY DEFAULTS<br />

To restore the factory default settings, press the ALARM LIMITS pushbutton to enter the Alarm<br />

Limits mode.<br />

Press and hold the ALARM LIMITS pushbutton for 2 seconds until the message “Saving? N”<br />

appears in the Message Window.<br />

Press the ALARM LIMITS pushbutton to bring up the Recall menu. The message “Recall? N”<br />

appears in the Message Window. Pressing either the ARROW UP or ARROR DOWN<br />

pushbutton will display the message “Recall? Y” in the Message Window.<br />

Press the ALARM LIMITS pushbutton. The Message Window will display "Restore" and return<br />

to the Alarm Limit menu. When finished, press the CANCEL pushbutton to return to the “Ready”<br />

state.<br />