CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />
The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> can automatically take a series of blood pressure measurements for<br />
a five (5) minute interval with a brief (approx. ten (10) second) pause between determinations to<br />
allow venous blood return.<br />
Press and Hold the START/STAT pushbutton for two (2) seconds to activate this feature. The<br />
Message Window will display "STAT XXX", where “XXX” is the real time cuff pressure. Between<br />
readings the Message Window will display "0:XX S" where “XX” is the 10-second count down<br />
until the next measurement.<br />
After five (5) minutes of determinations the monitor will stop taking measurements and exit the<br />
STAT mode. The monitor will return to the Manual mode and the Message Window will return to<br />
the "Ready" message.<br />
In the STAT Mode, while during a measurement or in the idle time between measurements,<br />
pressing the CANCEL pushbutton will cause the monitor to exit the STAT mode and return to<br />
the Manual mode. The monitor’s front panel NIBP numeric displays will indicate the reading of<br />
the previous measurement taken and the Message Window will display the time stamp of that<br />
reading.<br />
NOTE:<br />
Pressing the CYCLE pushbutton during the Manual, Automatic or STAT Modes will terminate the<br />
current NIBP measurement (if active) and place the monitor in the Automatic Cycle mode. The<br />
user will then be allowed to alter the Cycle time, if necessary, and begin a new NIBP<br />
measurement.<br />
WARNING:<br />
When monitoring over an extended period of time, or at frequent intervals, periodically observe<br />
the patient’s limb to make sure that the circulation is not impaired for a prolonged period of time.<br />
HISTORY MODES<br />
The History Modes allow the user to recall previously taken measurements. The measurement<br />
results and the time the measurement was taken are also displayed.<br />
Patient History is organized in two visually displayed lists.<br />
Event History: Pressing the HISTORY pushbutton the first time reveals the first viewable list that<br />
contains up to 480 entries of Event driven data (NIBP and Predictive Temperature (if installed))<br />
along with the corresponding %SpO 2 results (if installed).<br />
Trend History: Pressing the HISTORY pushbutton again reveals the second viewable list that<br />
contains up to 480 entries of Trend History at one (1) minute intervals. The Trend History<br />
contains all the events from the Event History log combined with readings that are saved every<br />
minute consisting of: a one (1) minute average of the %SpO 2 values (if installed) and <strong>Monitor</strong><br />
Mode Temperature values (if installed).<br />
88<br />
21-02-0298 REV. 00 12/06