<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

AUTOMATIC CYCLE FOR BLOOD PRESSURE DETERMINATION<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> can automatically take blood pressure measurements at pre-selected<br />

time intervals.<br />

To choose a time interval, press the CYCLE TIME pushbutton. The Message Window displays<br />

the format “Auto XX”. Where “XX” is the time interval in minutes.<br />

Choose the time interval using one of the following methods:<br />

1) Press the CYCLE TIME pushbutton to advance through the available time intervals.<br />

2) Press the ARROW UP or ARROW DOWN pushbutton keys to advance forward or<br />

backwards through the available time intervals.<br />

Once a time interval has been selected, press the START pushbutton to begin the first<br />

measurement.<br />

Between each measurement, the Message Window will display the time remaining until the next<br />

measurement, as well as the cycle time chosen in the format "MM:SS CC" where “MM” is the<br />

minutes and “SS” is the seconds until the next measurement and “CC” is the cycle time selected.<br />

The measurement results are displayed on the front panel until the start of the next<br />

measurement cycle.<br />

NOTE:<br />

If a measurement is desired between measurement cycles, press the START pushbutton. After<br />

this measurement, the monitor will re-enter the Automatic Cycle mode and countdown to the<br />

next measurement based on the Cycle time selected.<br />

In the Automatic Mode, while during a measurement or in the idle time between measurements,<br />

pressing the CANCEL pushbutton will cause the monitor to exit the Automatic mode and return<br />

to the Manual mode. The monitor’s front panel NIBP numeric displays will indicate the reading<br />

of the previous measurement taken and the Message Window will display the time stamp of that<br />

reading.<br />

NOTE:<br />

The Automatic Cycle mode is not saved during a power cycle.<br />

STAT MODE<br />

WARNING:<br />

Readings obtained during STAT mode may not meet the stated accuracy of this monitor.<br />

WARNING:<br />

In some cases, rapid, prolonged cycling of an oscillometric, noninvasive blood pressure monitor<br />

cuff has been associated with any or all of the following: ischemia, purpura, or neuropathy.<br />

Apply the oscillometric cuff appropriately, according to instructions, and check the cuff site and<br />

cuffed extremity regularly when blood pressure is measured at frequent intervals or over<br />

extended periods of time.<br />

21-02-0298 REV. 00 12/06<br />