<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

IMPORTANT:<br />

Excessive patient motion can contribute to inaccurate measurements. It is important that the<br />

patient be kept still during a measurement. Make every attempt to alleviate fear, anxiety and<br />

pain.<br />

Press the START pushbutton to begin a measurement. For the first measurement the monitor<br />

will inflate to the default setting or the Initial Inflation Pressure selected.<br />

NOTE:<br />

When the monitor is configured in the Alarms Off mode, the message “Alrm Off” is displayed in<br />

the Message Window, the cuff inflation pressure will always be set to 150 mmHg (Adult) and 85<br />

mmHg (Neonate). Also, as an example, if the Initial Inflation Pressure was set to 100 mmHg,<br />

the first NIBP inflation pressure would be 100 mmHg followed by pressures at 150 mmHg.<br />

The monitor’s front panel NIBP displays will indicate all dashes “- -“ while the measurement is in<br />

progress and the Message Window displays the Inflation Pressure in the format “Meas XXX”,<br />

where “XXX” is the pressure value. For subsequent measurements, the monitor will inflate<br />

approximately 30 mmHg higher than the previously determined Systolic pressure.<br />

The measurement typically takes less than 30 seconds to complete. In no case will the cuff<br />

remain pressurized for more than 120 seconds for Adult/Pediatric patients and no more than 90<br />

seconds for Neonates.<br />

When the measurement is completed, the cuff will automatically deflate and the monitor will<br />

display the Systolic, Diastolic, MAP (if enabled), and Pulse Rate values on the front panel<br />

displays. The Message Window displays a time stamp of the NIBP measurement in the format<br />

“BP HH:MM”. Where “HH” is the hour and “MM” is the minute of the measurement taken.<br />

NOTE:<br />

If SpO 2 is active, the Pulse Rate value will be determined from the SpO 2 measurement.<br />

Press the CANCEL pushbutton, at any time, to stop a measurement and deflate the cuff during<br />

the measurement process. The monitor’s front panel NIBP displays will indicate all dashes “- - “<br />

and the Message Window returns to “Ready”.<br />

The CANCEL pushbutton can be pressed after a measurement is taken to clear the current<br />

reading from the displays. The monitor’s front panel NIBP displays will indicate all dashes “- -“<br />

and the Message Window returns to “Ready”.<br />

NOTE:<br />

If any displayed NIBP measurement were to be left on the display for up to twenty-four (24)<br />

hours, the monitor will automatically blank the displays to all dashes “- -“ and the Message<br />

Window returns to “Ready”.<br />

86<br />

21-02-0298 REV. 00 12/06