<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

7. TEMPERATURE MONITORING<br />

WELCH ALLYN TEMPERATURE<br />

WARNING:<br />

During use, single-use disposable temperature probe covers supplied by CAS Medical Systems<br />

or Welch Allyn will limit patient cross-contamination. The use of any other probe cover or the<br />

failure to use a probe cover may produce temperature errors and will invalidate the monitor’s<br />

warranty. Temperature probe covers are required to ensure the safety of the patient and user.<br />

CAUTION:<br />

When replacing a SureTemp Plus probe, it is recommended to replace the Probe Well received<br />

with the probe. Failure to replace the Probe Well could result in patient cross contamination.<br />

TAKING AN ORAL TEMPERATURE<br />

CAUTION:<br />

Never use the Temperature Probe without a probe cover. Accurate oral temperatures can only<br />

be obtained by using the blue temperature probe. The use of the wrong probe will produce<br />

temperature errors.<br />

To take an oral temperature (in either Predictive or <strong>Monitor</strong> mode) follow this procedure:<br />

1) Insure that the oral probe is connected to the monitor and that the probe is secured into<br />

its holder. An audible “click” should be heard when the SureTemp® probe is completely<br />

placed into its holder. The SureTemp Plus® probes will slide in and out of the Probe<br />

Well without any restrictions. The oral probe has a blue probe cover ejection button.<br />

2) Remove the probe from the probe holder. A short self-test mode will be initiated where<br />

every LED segment in the TEMP display window are illuminated briefly. Following this<br />

self-test the display will show “OrL”. The “OrL” in the TEMP display window indicates that<br />

the oral algorithm will be used to take a predictive temperature measurement.<br />

3) Load a probe cover onto the probe by holding the probe collar with the thumb and<br />

forefinger, being careful not to hold or press the ejection button. Refer to Figure 8.<br />

Figure 8: Loading the Probe Cover<br />

21-02-0298 REV. 00 12/06<br />