<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

NONIN® OXIMETER (if so equipped)<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> can be equipped to use SpO 2 sensors manufactured by Nonin. No<br />

other manufacturer’s sensors should be used.<br />

NOTE:<br />

<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s equipped with Nonin oximetry will have the Nonin logo next to the SpO 2<br />

connector.<br />

ATTACHING THE SENSOR CABLE<br />

1) Select and apply a sensor that is the appropriate size for the patient’s digit or extremity,<br />

according to the instructions provided by Nonin.<br />

2) Plug the sensor assembly into the SpO 2 connector on the side panel of the monitor. The<br />

connector is shaped like a “D”. Line up the “D” on the sensor cable with the “D” on the<br />

receptacle. Push the connector in completely.<br />

3) Press the ON/STANDBY pushbutton to turn “ON” the monitor.<br />

4) Check the Alarm Limits and configure them appropriately for the patient. Refer to Page<br />

91, PATIENT ALARM MODE for more information.<br />

REMOVING THE SENSOR CABLE<br />

When SpO 2 monitoring is not required, disconnect the Sensor Cable by carefully removing the<br />

connector from the SpO 2 input receptacle.<br />

NOTE:<br />

To avoid damage to the Sensor Cable, always hold it by the connector rather than the cable<br />

when connecting or disconnecting either end.<br />

When the probe is disconnected from the monitor, the message “No Probe” appears in the<br />

Message Window and an audible alarm sounds indicating a connection has been broken. Press<br />

the SILENCE/RESET pushbutton to silence the visual and audible alarm.<br />

NOTE:<br />

If either the 2-Minute Audio Off or Permanent Audio Off is enabled, no audio will be heard but a<br />

visual message will appear in the Message Window.<br />

Refer to Section 13, ACCESSORIES for Nonin oximeter probe types and part number<br />

information. Consult instructions enclosed with each sensor for proper application.<br />