27.01.2014 Views

CASMED 740 Vital Signs Monitor - Implox

CASMED 740 Vital Signs Monitor - Implox

CASMED 740 Vital Signs Monitor - Implox

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

NOTE:<br />

For best results, secure the sensor cable independently from the sensor, preferably around the<br />

base of the finger. Tape may be used to secure the cable to the patient. Make sure that the<br />

tape being used does not restrict the blood flow.<br />

2) Once the sensor has been attached to the patient and to the monitor, the SpO 2 signal bar<br />

graph will illuminate and indicate the relative signal strength and signal quality at the<br />

sensor site. Wait for the monitor to determine the initial %SpO 2 and Pulse Rate values.<br />

When the values have been determined, they will be displayed in their respective %SpO 2<br />

and ♥BPM front panel LED display windows.<br />

NOTE:<br />

Inspect the pulse oximeter site every 2 to 4 hours or per hospital protocol. If there is any skin<br />

irritation caused by the sensor, remove the sensor and apply it to a different location.<br />

3) When SpO 2 monitoring is completed, remove the sensor from the patient.<br />

When the probe is removed from the patient, the message “Prb OFF” appears in the Message<br />

Window and an audible alarm sounds, indicating a connection has been lost.<br />

Press the SILENCE/RESET pushbutton. The monitor silences the audible alarm tone, but the<br />

message remains.<br />

NOTE:<br />

If either the 2-Minute Audio Off or Permanent Audio Off is enabled, no audio will be heard but a<br />

visual message will appear in the Message Window.<br />

MASIMO® OXIMETER (if so equipped)<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> can be equipped to use SpO 2 sensors manufactured by Masimo. No<br />

other manufacturer’s sensors should be used.<br />

NOTE:<br />

<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s equipped with Masimo oximetry will have the Masimo SET® logo next to<br />

the SpO 2 connector.<br />

ATTACHING THE CABLES<br />

1) Select and apply a sensor that is the appropriate size for the patient’s digit or extremity,<br />

according to the instructions provided by Masimo.<br />

2) Orient the connecting tab of the sensor so that the shiny contacts are pointed up. Mate<br />

the logo on the sensor to the logo on the patient cable.<br />

3) Insert the tab of the sensor into the patient cable connector until there is a tactile or<br />

audible “click” connection. Verify a secure connection and gently tug on the patient cable<br />

connector.<br />

4) Plug the Interface Cable into the SpO 2 connector on the side panel of the monitor. The<br />

connector is shaped like a “D”. Line up the “D” on the Interface Cable with the “D” on the<br />

receptacle. Push the cable in until you hear an audible “click”.<br />

21-02-0298 REV. 00 12/06<br />

55

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!