6. PULSE OXIMETRY MONITORING<br />
The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> features a wide variety of SpO 2 technology that is ideal for every<br />
application.<br />
TAKING A SpO 2 MEASUREMENT<br />
IMPORTANT:<br />
Prior to patient monitoring, ensure the monitor is configured to the appropriate patient mode –<br />
Neonate or Adult.<br />
The following is a general procedure for taking a SpO 2 measurement:<br />
1) Select a sensor based on the patient size and monitoring conditions and properly attach<br />
the sensor to the patient.<br />
FINGER CLIP SENSORS<br />
The finger clip sensor is designed for spot check monitoring of pediatric and adult patients or<br />
continuous monitoring where patient movement is not expected, and the patient’s finger is large<br />
enough for the sensor to be attached securely.<br />
NOTE:<br />
If patient movement is occurring or the finger size is inappropriate, select a different sensor that<br />
is appropriate for the patient and the monitoring environment.<br />
Figure 4: SpO 2 Finger Clip Sensor Application<br />
Insert the finger (preferably left or right index finger) completely into the sensor. Place the<br />
sensor with the LEDs positioned on the nail side. The thumb is specifically not recommended<br />
for use with the finger clip sensor.<br />
21-02-0298 REV. 00 12/06<br />