27.01.2014 Views

CASMED 740 Vital Signs Monitor - Implox

CASMED 740 Vital Signs Monitor - Implox

CASMED 740 Vital Signs Monitor - Implox

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

Measurements made above the level of the heart will give reduced blood pressure readings<br />

while measurements made below the heart level will give increased readings. These errors are<br />

mainly due to the weight of the blood.<br />

The following guidelines should be observed;<br />

• When applying the cuff, make sure the cuff tubing is centered over the brachial artery.<br />

• Wrap the cuff for a snug fit to avoid prolonged pumping time. Do not wrap a cuff over the<br />

patient’s clothing.<br />

• Verify the hook and loop sections of the cuff are fully engaged when it is wrapped around<br />

the limb.<br />

• The limb should be positioned to be at heart level.<br />

• Do not compress the cuff or the cuff hose. The hose must not be kinked or pinched.<br />

Figure 3: Cuff Positioning<br />

48<br />

21-02-0298 REV. 00 12/06