27.01.2014 Views

CASMED 740 Vital Signs Monitor - Implox

CASMED 740 Vital Signs Monitor - Implox

CASMED 740 Vital Signs Monitor - Implox

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

5. BLOOD PRESSURE MONITORING<br />

CUFF SELECTION AND APPLICATION<br />

The use of properly designed and sized cuffs is essential for the accurate measurement of blood<br />

pressure. SoftCheck® and Safe-Cuff TM single patient use cuffs or Tuff-Cuff® and UltraCheck®<br />

reusable cuffs are recommended for use with the <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>.<br />

NOTE:<br />

CAS recommends the use of its reusable, disposable and neonatal cuffs with the <strong>CASMED</strong> <strong>740</strong><br />

<strong>Monitor</strong>.<br />

CAS recommends the use of the inflation hose supplied with the monitor or a replacement from<br />

CAS.<br />

The widest cuff that can be placed around the upper arm or thigh should be used. A cuff that is<br />

too small for the arm will not supply sufficient pressure to the artery. This can cause an<br />

erroneously high blood pressure reading. Substitution of a cuff different from that supplied might<br />

result in a measurement error.<br />

NOTE:<br />

Overlapping the cuff will not affect the measurement results.<br />

Figure 2: Cuff Application<br />

As an example, the end of the Tuff-Cuff is marked with a white arrow. To insure that the proper<br />

cuff size has been selected, wrap the cuff around the patient’s extremity, the marker should fall<br />

between the white markings on the cuff.<br />

21-02-0298 REV. 00 12/06<br />

47

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!