<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

GENERAL NOTES:<br />

There are no known risks with common disposal of equipment or accessories; however, the<br />

disposing of accessories should follow in accordance with local hospital policies. The user<br />

should ensure these policies do not conflict with any local, state or federal guidelines.<br />

The monitor is suitable for use in the presence of electro surgery.<br />

The monitor is suitable to be connected to public AC mains power.<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> is not “Category AP or APG Equipment”.<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> is for “Continuous Operation”.<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> applied parts are “Type BF Defibrillation Proof”.<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> provides “DRIP-PROOF” level of protection from ingress to moisture.<br />

Do not expose the <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> to extreme moisture levels such as direct exposure to<br />

rain. Exposure to extreme moisture levels may cause incorrect or inaccurate performance or<br />

device failure during or after exposure.<br />

AUTOMATIC SAFETY FEATURES<br />

The monitor has been designed to promote patient safety. The maximum amount of time<br />

allowed to complete a blood pressure measurement is 120 seconds in adult mode and 90<br />

seconds in neonate mode. If the measurement has not been completed within that time, the cuff<br />

is deflated automatically and a message is displayed indicating the problem.<br />

To prevent exposure of the extremity to an inordinately high pressure, the cuff is deflated<br />

automatically when the pressure in the system is greater than 290 mmHg in the adult mode or<br />

145 mmHg in the neonatal mode.<br />

The cuffs used by the <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> are designed without transducers for patient safety.<br />

The transducers used for NIBP measurement are located inside the monitor on the NIBP board<br />

and are isolated from the patient.<br />

In the event of a microprocessor failure, the cuff will be deflated automatically within ten (10)<br />

seconds.<br />

All equipment parts are protected against the effects of the discharge of a defibrillator. No<br />

separate actions are required when using this equipment with a defibrillator.<br />

Should the AC or DC power be interrupted coming into the monitor, the monitor automatically<br />

runs off battery power. An indication of this would be a change in color of the Battery Power<br />

Visual Indicator LED from Green to either Orange or Red.<br />

21-02-0298 REV. 00 12/06<br />