CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />
4. SAFETY MEASURES AND WARNINGS<br />
WARNING:<br />
Do not use this instrument for any purpose other than specified in this manual. Doing so will<br />
invalidate the monitor’s warranty.<br />
Do not connect more than one (1) patient to the monitor.<br />
Do not plug the monitor into an outlet controlled by a wall switch.<br />
Before each use, verify that the alarm limits are appropriate for the patient being monitored.<br />
The position of subject, physiological condition and other factors affect the readings.<br />
Blood pressure and pulse can fluctuate greatly between measurements; the monitor cannot alert<br />
the user to changes in vital signs occurring between measurement cycles.<br />
Occasionally, electrical signals at the heart do not produce a peripheral pulse. If a patient’s<br />
beat-to-beat pulse amplitude varies significantly (for example, pulsus alternans, atrial fibrillation,<br />
rapid-cycling artificial ventilator), blood pressure and pulse rate readings can be erratic and an<br />
alternate measuring method should be used for confirmation.<br />
Where the integrity of the external protective conductor in the installation or its arrangement is in<br />
doubt, EQUIPMENT shall be operated from its INTERNAL ELECTRICAL POWER SOURCE.<br />
Isolation of product from mains can only be achieved by removal of external power cord.<br />
Do not, under any circumstances, perform any testing or maintenance on the monitor or power<br />
cord while the unit is being used to monitor a patient. Unplug the power cord before cleaning or<br />
servicing the monitor. The operator should not perform any servicing except as specifically<br />
stated in this manual.<br />
Do not touch part of non-medical electrical equipment in the patient environment after removal of<br />
covers, connectors etc… without the use of a tool which operate at voltages not exceeding 25<br />
VAC or 60 VDC and the patient at the same time.<br />
Do not use a frayed or damaged power cord, or any accessory if you notice any sign of damage.<br />
Contact CAS Medical Systems for assistance.<br />
The use of Accessory equipment not complying with the equivalent safety requirements of this<br />
equipment may lead to a reduced level of safety of the resulting system. Consideration relating<br />
to the choice shall include:<br />
- Use of the accessory in the Patient Environment.<br />
- Evidence that the safety certification of the accessory has been performed in accordance<br />
with the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard.<br />
Equipment not suitable for use in the presence of FLAMMABLE ANESTHETICS.<br />
Equipment is not intended to be used in Oxygen Enriched Atmospheres.<br />
21-02-0298 REV. 00 12/06<br />
39