<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

1. INTRODUCTION AND INTENDED USE<br />

INTRODUCTION<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> is a multi parameter monitor measuring blood pressure, oxygen<br />

saturation and temperature. Non-invasive blood pressure is measured using the oscillometric<br />

technique determining systolic, diastolic, mean arterial pressure and pulse rate. The pulse<br />

oximeter function continuously monitors and displays values for functional arterial hemoglobin<br />

saturation and a pulse rate. Temperature is obtained in the normal (predictive) mode in as little<br />

as four (4) seconds. A monitoring mode is available for taking axillary temperatures.<br />

INDICATIONS FOR USE<br />

The <strong>CASMED</strong> <strong>740</strong> Series <strong>Vital</strong> <strong>Signs</strong> <strong>Monitor</strong> is indicated for use for non-invasive monitoring of<br />

blood pressure, oxygen saturation, pulse and temperature of adult, pediatric and neonatal<br />

patients, in the care of health professionals.<br />

WARNING:<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> is intended only as an adjunct in patient assessment. It must be<br />

used in conjunction with clinical signs and symptoms.<br />

CONTRAINDICATIONS<br />

• Oral and Rectal Temperature measurements are not intended for neonatal use.<br />

• Reusable SpO 2 sensors are contraindicated for use for prolonged periods of use. They<br />

are not intended for long term monitoring. They must be removed and repositioned<br />

every four (4) hours and if indicated by circulatory condition or skin integrity, reapplied to<br />

a different monitoring site.<br />

• Disposable SpO 2 sensors are contraindicated for patients that exhibit allergic reactions to<br />

adhesive tape. The sensors must be removed and repositioned every eight (8) hours<br />

and if indicated by circulatory condition or skin integrity, reapplied to a different<br />

monitoring site.<br />

• No other contraindications are known at this time.<br />

BRIEF DEVICE DESCRIPTION<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> is compact, lightweight and portable, allowing it to be easily carried<br />

and used in a variety of clinical settings. The monitor is powered by AC Line Power, +12 VDC or<br />

by a Nickel Metal Hydride (NiMH) rechargeable battery pack. The internal battery pack charges<br />

when the monitor is plugged into a power source (AC Line Power or +12 VDC). The <strong>CASMED</strong><br />

<strong>740</strong> <strong>Monitor</strong> can be set to operate in one of nine (9) different languages: English, German,<br />

French, Italian, Spanish, Dutch, Swedish, Portuguese or Norwegian. The message window can<br />

display various system alarm messages. These messages direct the user to check conditions<br />

such as the battery state, air leaks and measurement problems. The message window also<br />

displays the operational mode of the monitor (automatic or manual).<br />

21-02-0298 REV. 00 12/06<br />