<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

TEMPERATURE (OPTIONAL)<br />

Characteristic<br />

Temperature Range:<br />

Accuracy:<br />

28.9 to 42.2 °C (84.0 to 108.0 °F)<br />

Specification<br />

+/-0.1°C (+/-0.2°F), Meets or exceeds ASTM Standards<br />

PATIENT ALARMS<br />

<strong>CASMED</strong> <strong>740</strong><br />

with NIBP<br />

Patient Parameter Neonatal Limit Range Adult Limit Range<br />

Low High Low High<br />

SYS 35 – 130 35 – 130 35 – 250 35 – 250<br />

DIA 20 – 105 20 – 105 20 – 215 20 – 215<br />

MAP 25 – 120 25 – 120 25 – 230 25 – 230<br />

Pulse 45 – 235 45 – 235 35 – 235 35 – 235<br />

<strong>CASMED</strong> <strong>740</strong><br />

with NIBP and<br />

SpO 2<br />

Patient Parameter Neonatal Limit Range Adult Limit Range<br />

Low High Low High<br />

SYS 35 – 130 35 – 130 35 – 250 35 – 250<br />

DIA 20 – 105 20 – 105 20 – 215 20 – 215<br />

MAP 25 – 120 25 – 120 25 – 230 25 – 230<br />

%SpO 2 70 – 95 80 – 99 70 – 95 80 – 99<br />

Pulse 25 – 235 25 – 235 25 – 235 25 – 235<br />

NOTE:<br />

Each alarm limit may also be selected “OFF” individually or as a whole.<br />

Low Limits cannot be set above the associated High Limit.<br />

High Limits cannot be set lower than the associated Low Limit.<br />

CONTROL PANEL<br />

Characteristic<br />

Display:<br />

Parameters Displayed:<br />

Specification<br />

LED display of measurement results, instructions, troubleshooting messages and<br />

signal strength bar.<br />

Systolic Pressure, Diastolic Pressure and Mean Arterial Pressure (MAP)<br />

Pulse Rate<br />

%SpO 2<br />

Temperature (in Fahrenheit or Celsius)<br />

SAFETY LIMITS<br />

Characteristic<br />

Automatic Cuff Deflation:<br />

Specification<br />

If cuff pressure exceeds 290 mmHg (Adult); 145 mmHg (Neonate)<br />

If measurement time exceeds 120 seconds (Adult), 90 seconds (Neonate)<br />

If safety timer detects microprocessor failure<br />

164<br />