<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

Characteristic<br />

Specification<br />

Nellcor® OxiMax®<br />

Type:<br />

Functional Oxygen Saturation<br />

SpO 2 % Range: 1 - 100%<br />

SpO 2 Accuracy: Sensor Accuracy<br />

MAX-A<br />

70 - 100%, +/-2 digits (1 S.D.)<br />

MAX-AL<br />

MAX-N (Adult)<br />

MAX-P<br />

MAX-I<br />

MAX-FAST<br />

SC-A (Adult)<br />

Sensor<br />

Accuracy<br />

OxiCliq A<br />

70 - 100%, +/-2.5 digits (1 S.D.)<br />

OxiCliq I<br />

OxiCliq N (Adult)<br />

OxiCliq P<br />

Sensor<br />

Accuracy<br />

DS-100A<br />

70 - 100%, +/-3 digits (1 S.D.)<br />

D-YS (Infant to Adult)<br />

MAX-N (Neonate)<br />

OXI-A/N (Adult)<br />

OXI-P/I<br />

SC-NEO (Neonate)<br />

SC-PR (Neonate)<br />

Sensor<br />

Accuracy<br />

D-YS<br />

70 - 100%, +/-3.5 digits (1 S.D.)<br />

D-YSE<br />

D-YSPD<br />

OxiCliq N (Neonate)<br />

Sensor<br />

Accuracy<br />

MAX-R<br />

80 - 100%, +/-3.5 digits (1 S.D.)<br />

Sensor<br />

Accuracy<br />

D-YS (Neonate) 70 - 100%, +/-4.0 digits (1 S.D.)<br />

OXI-A/N (Neonate)<br />

Measurement Wavelengths:<br />

Power:<br />

Pulse Rate Range:<br />

Pulse Rate Accuracy:<br />

Numerics:<br />

Red 660 Nanometers<br />

Infrared 890 Nanometers<br />

Not exceeding 15 mW<br />

20 - 240 BPM<br />

+/-3 digits<br />

Updated every one (1) second.<br />

NOTE:<br />

For further information on sensors and sensor accuracy, contact Nellcor.<br />