<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

Characteristic<br />

Specification<br />

Masimo SET®<br />

Type:<br />

Functional Oxygen Saturation<br />

SpO 2 % Range: 0 - 100%<br />

SpO 2 Accuracy: Sensor Accuracy<br />

DC-195<br />

70 - 100%, +/-2 digits (1 S.D.)<br />

LNOP® Adt<br />

LNOP DCI<br />

LNOP DCSC<br />

LNOP DC1P<br />

LNOP DC150<br />

LNOP Pdt<br />

LNOP Y-1<br />

LNOPv Ad<br />

Sensor<br />

Accuracy<br />

LNOP Neo<br />

70 - 100%, +/-3 digits (1 S.D.)<br />

LNOP Neo Pt<br />

Sensor<br />

Accuracy<br />

LNOP EAR<br />

70 - 100%, +/-3.5 digits (1 S.D.)<br />

Measurement Wavelengths:<br />

Power:<br />

Pulse Rate Range:<br />

Pulse Rate Accuracy:<br />

Numerics:<br />

Red 660 Nanometers<br />

Infrared 905 Nanometers<br />

Maximum radiant power at 50 mA pulsed is 0.79mW<br />

25 - 240 BPM<br />

+/-3 BPM<br />

Updated every one (1) second.<br />

NOTE:<br />