CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />
15. SPECIFICATIONS<br />
NIBP MEASUREMENT<br />
Characteristic<br />
Technique:<br />
Patient Range:<br />
Specification<br />
Oscillometric (MAX NIBP® Technology)<br />
Microprocessor software eliminates most ambient noise and motion artifact.<br />
Neonate – Adult<br />
Blood Pressure Range NEO ADULT<br />
Systolic: 30 – 135 mmHg 30 – 255 mmHg<br />
Diastolic: 15 – 110 mmHg 15 – 220 mmHg<br />
MAP: 20 – 125 mmHg 20 – 235 mmHg<br />
Pulse Rate Range: 40 – 240 BPM 30 – 240 BPM<br />
Accuracy<br />
Blood Pressure:<br />
Pulse Rate:<br />
+/-5 mmHg with a standard deviation no greater than 8 mmHg (See Standards)<br />
+/-2% or +/-2 BPM, whichever is greater<br />
OXIMETRY (OPTIONS)<br />
Characteristic<br />
Specification<br />
Nonin®<br />
Type:<br />
Functional Oxygen Saturation<br />
SpO 2 % Range: 0 - 100%<br />
SpO 2 Accuracy: Sensor Accuracy<br />
8000AA<br />
70 - 100%, +/-2 digits (1 S.D.)<br />
8000AP<br />
8000K2<br />
Sensor<br />
Accuracy<br />
8000J<br />
70 - 100%, +/-3 digits (1 S.D.)<br />
8000R<br />
Sensor<br />
Accuracy<br />
8000Q<br />
70 - 100%, +/-4 digits (1 S.D.)<br />
Measurement Wavelengths:<br />
Power:<br />
Pulse Rate Range:<br />
Pulse Rate Accuracy:<br />
Numerics:<br />
Red 660 nanometers<br />
Infrared 910 Nanometers<br />
3 mW nominal<br />
18 - 240 BPM<br />
+/-3% or +/-1 digit, whichever is greater<br />
Updated every one (1) second.<br />
NOTE:<br />
For further information on sensors and sensor accuracy, contact Nonin.<br />
21-02-0298 REV. 00 12/06<br />
161