<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

OXIMETRY CALIBRATION CHECK<br />

The oximeter is factory calibrated to determine the percentage of arterial oxygen saturation of<br />

functional hemoglobin. No user calibration is required.<br />

REPLACING THE MONITOR BATTERY<br />

A part number for the battery can be found in the Accessories section of this manual or on the<br />

label located on the inside panel of the battery pack. When the battery fails to hold a charge it<br />

will need to be replaced.<br />

CAS Medical Systems recommends the battery be changed every two (2) years.<br />

REMOVING THE BATTERY<br />

1) Turn the monitor “OFF” and disconnect the power cord.<br />

2) Push down on the battery latch to unlock the battery door from the rear panel of the<br />

monitor.<br />

3) Carefully remove the battery pack from the rear panel of the monitor. Refer to Figure 25.<br />

Figure 25: Removing the <strong>Monitor</strong> Battery Pack<br />

140<br />

21-02-0298 REV. 00 12/06