<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

TEMPERATURE CALIBRATION CHECK<br />

Verify the calibration of the monitor’s Temperature circuit, once (1) every year.<br />

To perform a Temperature Calibration Check, obtain a Calibration Key. This key can be<br />

purchased directly from Welch Allyn or CAS Medical Systems. Refer to Section 13,<br />

ACCESSORIES for part number information.<br />

NOTE:<br />

The Temperature Calibration Check can be performed, at any time, once the monitor enters the<br />

Configuration Menu.<br />

1) Turn the monitor “OFF”.<br />

2) Remove the temperature probe and its connector completely from the monitor and insert<br />

the Calibration Key.<br />

3) Turn the monitor “ON”, and enter the monitor’s Configuration Menu. Refer to Page 135,<br />

ENTER THE TEST MODE.<br />

4) Re-insert and remove the temperature probe from the probe guide to reset the<br />

thermometer’s electronics. An audible “click” should be heard when the SureTemp®<br />

probe is completely placed into its holder. The SureTemp Plus® probes will slide in and<br />

out of the Probe Well without any restrictions.<br />

5) Wait for the test to complete, and observe the display reading in the TEMP display<br />

window.<br />

6) The display value should read 36.3 +/- 0.1 °C or 97.3 +/-0.1°F.<br />

NOTE:<br />

The monitor will display the Temperature Calibration Key value using the current temperature<br />

units selected.<br />

NOTE:<br />

The Temperature Calibration Key will only operate while the monitor is in the Configuration<br />

Menu. If the Calibration Key is inserted during normal monitoring, the TEMP display will show<br />

three (3) flashing dashes “- - -“ and no value will be displayed.<br />

NOTE:<br />

While in the Configuration Menu, if no pushbutton is pressed within 60 seconds, the monitor will<br />

automatically exit the <strong>Monitor</strong> Configuration menu. The Message Window will briefly display<br />

“Saving” and return to the “Ready” mode.<br />

7) Remove the Calibration Key and re-insert the temperature probe connector.<br />

8) Install the temperature probe into the probe holder. An audible “click” should be heard<br />

when the SureTemp® probe is completely placed into its holder. The SureTemp Plus®<br />

probes will slide in and out of the Probe Well without any restrictions.<br />

21-02-0298 REV. 00 12/06<br />

139