5) At the completion of a successful Pressure Check, the Message Window will display<br />

“Passed”, the monitor will beep two (2) times and will return to the Calibration Check<br />

function after five (5) seconds.<br />

6) If the monitor fails the Pressure Check, the Message Window will display "Leak", the<br />

monitor will beep three (3) times and the return to the Calibration Check function after<br />

five (5) seconds.<br />

Due to the volume differences of the hoses offered with the <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>, the monitor<br />

may incorrectly fail the Plug Tube check. Should the monitor fail the Plug Tube Pressure Check,<br />

obtain a 500 ml Pressure Cylinder and follow the 500 ml Pressure Check.<br />

500 ml PRESSURE CHECK<br />

Obtain a fixed volume 500 ml Pressure Cylinder (CAS p/n 01-02-0248).<br />

1) Place the end of the monitor’s inflation hose securely onto the luer fitting at the top of the<br />

pressure cylinder. The hose must fit securely onto the connector for this test to be<br />

performed properly.<br />

2) Enter the Test Mode. Refer to Page 135, ENTER THE TEST MODE. The Message<br />

Window will briefly display “TestMode” followed by "0 mmHg".<br />

3) Press the NIBP START pushbutton to begin the Pressure Check.<br />

4) The Message Window will display “Chk Prs”, will inflate to approximately 160 mmHg and<br />

attempt to hold this pressure. The pressure value will be displayed in the SYSTOLIC<br />

display window. This test takes about fifteen (15) seconds.<br />

5) At the completion of a successful Pressure Check, the Message Window will display<br />

“Passed”, the monitor will beep two (2) times and will return to the Calibration Check<br />

function after five (5) seconds.<br />

6) If the monitor fails the Pressure Check, the Message Window will display "Leak", the<br />

monitor will beep three (3) times and the return to the Calibration Check function after<br />

five (5) seconds.<br />

Should the monitor fail the 500 ml Pressure Check, it is recommended the monitor be returned<br />

to CAS Medical Systems for service.<br />