<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

0 mmHg +/- 1 mmHg<br />

50 mmHg +/- 4 mmHg<br />

100 mmHg +/- 4 mmHg<br />

150 mmHg +/- 4 mmHg<br />

200 mmHg +/- 5 mmHg<br />

NOTE:<br />

If the monitor does not display the test pressure for the 30-second period, deflate to zero and<br />

verify the proper assembly of the calibration set-up. Re-inflate the system. If the monitor again<br />

fails to hold the pressure, it is recommended the monitor be returned to CAS Medical Systems<br />

for service.<br />

OVERPRESSURE<br />

Inflate the pressure slowly until 290 mmHg +/- 10 mmHg is reached. The Message Window<br />

should stop updating, display the message "OverPres" and provide a NIBP Application Error<br />

audible tone.<br />

Press the CANCEL pushbutton to exit the Overpressure Test. The monitor returns to the<br />

Calibration Check function.<br />

If the monitor does not meet the above specifications, it is recommended the monitor be<br />

returned to CAS Medical Systems for service.<br />

PNEUMATIC PRESSURE CHECKS<br />

Check the monitor's pneumatic system for air leakage every six (6) months.<br />

PLUG TUBE<br />

Obtain the male luer plug found in the Calibration Kit (product #P9) supplied with the monitor.<br />

1) Place this plug into the cuff connector at the end of the monitor inflation hose and twist<br />

one-quarter turn. The plug must fit securely into the connector for this test to be<br />

performed properly.<br />

2) Enter the Test Mode. Refer to Page 135, ENTER THE TEST MODE. The Message<br />

Window will briefly display “TestMode” followed by "0 mmHg".<br />

3) Press the NIBP START pushbutton to begin the Pressure Check.<br />

4) The Message Window will display “Chk Prs”, will inflate to approximately 180 mmHg and<br />

attempt to hold this pressure. The pressure value will be displayed in the SYSTOLIC<br />

display window. This test takes about fifteen (15) seconds.<br />

21-02-0298 REV. 00 12/06<br />

137