<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

EXIT THE TEST MODE<br />

When you have completed with the Test Mode, press the CANCEL pushbutton to exit. The<br />

Message Window will briefly display “Saving” and return to the “Ready” state.<br />

LED CHECK<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> incorporates an all lights “ON” check to verify the functionality of the<br />

front panel LED displays and indicators.<br />

Enter the Test Mode. Refer to Page 135, ENTER THE TEST MODE. The Message Window<br />

will briefly display “TestMode” followed by "0 mmHg".<br />

Press and hold either the ARROW UP or ARROW DOWN pushbuttons. The monitor will<br />

illuminate all appropriate 7-segment displays, bar graph, bell icon and patient mode indicators<br />

for as long as the button is pressed.<br />

CALIBRATION CHECK<br />

Verify the calibration of the monitor once (1) per year.<br />

A Calibration Kit, (product #P9) is included with the monitor. The kit contains a T-connector with<br />

a male and a female luer fitting (for a Calibration Check) and a male luer plug (to be used for the<br />

Pneumatic Check).<br />

Obtain a mercury manometer whose accuracy meets the AAMI/ANSI Standard for Non-<br />

Automated Sphygmomanometers, 2002.<br />

NOTE:<br />

<strong>Monitor</strong> must be in the Adult mode prior to performing these pressure checks.<br />

SYSTEM PRESSURE<br />

Assemble the Calibration Kit according to the diagram provided in the P9 kit.<br />

1) Remove the manometer tubing from the inflation bulb. Connect the open ended tubing<br />

of the T-connector to the inflation bulb.<br />

2) Connect the female luer fitting to the inflation tube leading to the manometer.<br />

3) Connect the male luer fitting to the manometer tubing.<br />

4) Enter the Test Mode. Refer to Page 135, ENTER THE TEST MODE. The Message<br />

Window will briefly display “TestMode” followed by "0 mmHg".<br />

5) Use the manometer inflation bulb to slowly inflate the system pausing for 30 seconds at<br />

the following points and verify calibration according to the following table:<br />

136<br />

21-02-0298 REV. 00 12/06