EXIT THE TEST MODE<br />
When you have completed with the Test Mode, press the CANCEL pushbutton to exit. The<br />
Message Window will briefly display “Saving” and return to the “Ready” state.<br />
LED CHECK<br />
The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> incorporates an all lights “ON” check to verify the functionality of the<br />
front panel LED displays and indicators.<br />
Enter the Test Mode. Refer to Page 135, ENTER THE TEST MODE. The Message Window<br />
will briefly display “TestMode” followed by "0 mmHg".<br />
Press and hold either the ARROW UP or ARROW DOWN pushbuttons. The monitor will<br />
illuminate all appropriate 7-segment displays, bar graph, bell icon and patient mode indicators<br />
for as long as the button is pressed.<br />
CALIBRATION CHECK<br />
Verify the calibration of the monitor once (1) per year.<br />
A Calibration Kit, (product #P9) is included with the monitor. The kit contains a T-connector with<br />
a male and a female luer fitting (for a Calibration Check) and a male luer plug (to be used for the<br />
Pneumatic Check).<br />
Obtain a mercury manometer whose accuracy meets the AAMI/ANSI Standard for Non-<br />
Automated Sphygmomanometers, 2002.<br />
NOTE:<br />
<strong>Monitor</strong> must be in the Adult mode prior to performing these pressure checks.<br />
SYSTEM PRESSURE<br />
Assemble the Calibration Kit according to the diagram provided in the P9 kit.<br />
1) Remove the manometer tubing from the inflation bulb. Connect the open ended tubing<br />
of the T-connector to the inflation bulb.<br />
2) Connect the female luer fitting to the inflation tube leading to the manometer.<br />
3) Connect the male luer fitting to the manometer tubing.<br />
4) Enter the Test Mode. Refer to Page 135, ENTER THE TEST MODE. The Message<br />
Window will briefly display “TestMode” followed by "0 mmHg".<br />
5) Use the manometer inflation bulb to slowly inflate the system pausing for 30 seconds at<br />
the following points and verify calibration according to the following table:<br />