11. MAINTENANCE<br />

MAINTENANCE INTERVALS<br />

Preventive maintenance of the monitor is an important function that should be performed<br />

routinely by the user to ensure safe and efficient monitor operation. CAS Medical Systems<br />

recommends that you do the following:<br />

• Perform a pneumatic check every six (6) months.<br />

• Perform a calibration check once per year or when there is doubt about the validity of the<br />

pressure readings.<br />

• Replace the battery pack every two (2) years.<br />

If the monitor is in need of repair, it must be referred to the appropriate service personnel.<br />

Service performed by unauthorized personnel could be detrimental to the monitor and may void<br />

the warranty. For service, contact your dealer or CAS Medical Systems, Inc.<br />

TEST MODE<br />

The monitor must be in the Configuration Menu in order to perform the following functions:<br />

• All LEDs “ON” Check<br />

• Calibration Check<br />

System Pressure<br />

Over Pressure<br />

• Pneumatic Pressure Checks<br />

• Temperature Calibration Check<br />

ENTER THE TEST MODE<br />

To enter the monitor’s Configuration Menu, press and hold the VOLUME<br />

and ALARM<br />

LIMITS pushbutton keys while the monitor is being turned “ON” .<br />

Once in the menu, press one of the PREVIOUS (CYCLE TIME / ALARM LIMITS) programmed<br />

pushbutton keys until the Message Window briefly displays “TestMode” followed by “0 mmHg”.<br />

NOTE:<br />

While in the Test Mode if no pushbutton is pressed within 15 minutes, the monitor will<br />

automatically terminate the <strong>Monitor</strong> Configuration menu and return to the “Ready” mode.<br />

WARNING:<br />

Do not place the monitor in the TEST MODE when a cuff is attached to a patient.<br />

21-02-0298 REV. 00 12/06<br />