<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

ERROR MESSAGES ON THE MESSAGE WINDOW (cont.)<br />

ERROR<br />

MESSAGE<br />

"RangeErr"<br />

“SetClock”<br />

“Signal ?”<br />

(Masimo)<br />

POSSIBLE CAUSE<br />

The systolic reading exceeds the<br />

measurement range of 255 mmHg<br />

in the Adult mode or 135 mmHg in<br />

the Neonatal mode.<br />

The monitor’s clock needs to be<br />

set.<br />

The quality of the signal level being<br />

received by the SpO 2 sensor is in<br />

question.<br />

Repeat measurement.<br />

POSSIBLE SOLUTION<br />

If the message is displayed again, use another<br />

method to measure the patient’s blood pressure.<br />

The monitor’s time and date values are incorrect.<br />

Refer to Page 104, for information to set the Time<br />

and Date.<br />

The monitor’s internal clock battery needs to be<br />

replaced. Contact CAS Medical Systems.<br />

Ensure proper sensor type and application.<br />

Verify that the sensor emitter and detector are<br />

parallel to and directly opposing each other.<br />

Clean or replace the sensor.<br />

“Sig Sat” Motion pulses too strong. Limit patient activity; the arm must be still and/or<br />

relaxed.<br />

Repeat measurement.<br />

“Sys Err”<br />

"Time Out"<br />

"Weak Sig"<br />

An electronic failure has occurred<br />

within the monitor.<br />

The monitor was unable to<br />

complete a measurement within<br />

120 seconds in the Adult mode or<br />

90 seconds in the Neonatal mode.<br />

The monitor did not detect any<br />

pulses during a NIBP<br />

measurement.<br />

Contact CAS Medical Systems to have the<br />

monitor serviced.<br />

An extremely long measurement can be due to a<br />

loose cuff, high blood pressure, or monitor repumps.<br />

Try measurement again. Try higher initial<br />

pressure.<br />

If message consistently reappears try using<br />

another means to obtain patient's blood pressure.<br />

Check the fit of the cuff.<br />

Repeat measurement.<br />

Table 8: Error Messages on the Message Window<br />

21-02-0298 REV. 00 12/06<br />