ERROR MESSAGES ON THE MESSAGE WINDOW (cont.)<br />
ERROR<br />
MESSAGE<br />
"Motion"<br />
"NBP Cal"<br />
"NIBP Err"<br />
“No Probe”<br />
POSSIBLE CAUSE<br />
There was too much extremity<br />
motion for the monitor to<br />
accurately complete the NIBP<br />
measurement in 120 seconds.<br />
Pressure calibration data corrupted<br />
within NIBP module.<br />
An electronic failure has occurred<br />
within the NIBP module.<br />
The monitor is not detecting the<br />
SpO 2 probe.<br />
POSSIBLE SOLUTION<br />
Measurements can be obtained when there is<br />
limited extremity movement, but the<br />
measurement time may be extended.<br />
Measurement time is limited to 120 seconds.<br />
Restrain patient extremity motion.<br />
Pressure module needs recalibration.<br />
Contact CAS Medical Systems to have the<br />
monitor serviced.<br />
Contact CAS Medical Systems to have the<br />
monitor serviced.<br />
The probe was disconnected from either the<br />
Interface Cable or from the monitor.<br />
"OverPres" Cuff pressure exceeded 290<br />
mmHg in the Adult mode or 145<br />
mmHg in the Neonatal mode.<br />
Very rapid squeezing of the cuff can cause this<br />
error.<br />
Repeat the measurement.<br />
If this message repeatedly occurs during normal<br />
use, the monitor must be serviced.<br />
“P Search” The monitor is searching for a<br />
Pulse signal.<br />
Normal at power-up as the monitor searches for a<br />
pulse.<br />
“Prb OFF”<br />
"Pwr Fail"<br />
The monitor is no longer receiving<br />
a patient signal from the SpO 2<br />
probe.<br />
Power was disconnected from the<br />
monitor.<br />
The probe position may have changed.<br />
Check the probe site.<br />
The probe is no longer in contact with the patient.<br />
Check the probe site.<br />
Press the SILENCE/RESET pushbutton to clear<br />
the message.<br />
OR<br />
Re-cycle the monitor’s power.<br />
Table 8: Error Messages on the Message Window<br />
114<br />
21-02-0298 REV. 00 12/06