ERROR MESSAGES ON THE MESSAGE WINDOW (cont.)<br />

ERROR<br />

MESSAGE<br />

"Motion"<br />

"NBP Cal"<br />

"NIBP Err"<br />

“No Probe”<br />

POSSIBLE CAUSE<br />

There was too much extremity<br />

motion for the monitor to<br />

accurately complete the NIBP<br />

measurement in 120 seconds.<br />

Pressure calibration data corrupted<br />

within NIBP module.<br />

An electronic failure has occurred<br />

within the NIBP module.<br />

The monitor is not detecting the<br />

SpO 2 probe.<br />

POSSIBLE SOLUTION<br />

Measurements can be obtained when there is<br />

limited extremity movement, but the<br />

measurement time may be extended.<br />

Measurement time is limited to 120 seconds.<br />

Restrain patient extremity motion.<br />

Pressure module needs recalibration.<br />

Contact CAS Medical Systems to have the<br />

monitor serviced.<br />

Contact CAS Medical Systems to have the<br />

monitor serviced.<br />

The probe was disconnected from either the<br />

Interface Cable or from the monitor.<br />

"OverPres" Cuff pressure exceeded 290<br />

mmHg in the Adult mode or 145<br />

mmHg in the Neonatal mode.<br />

Very rapid squeezing of the cuff can cause this<br />

error.<br />

Repeat the measurement.<br />

If this message repeatedly occurs during normal<br />

use, the monitor must be serviced.<br />

“P Search” The monitor is searching for a<br />

Pulse signal.<br />

Normal at power-up as the monitor searches for a<br />

pulse.<br />

“Prb OFF”<br />

"Pwr Fail"<br />

The monitor is no longer receiving<br />

a patient signal from the SpO 2<br />

probe.<br />

Power was disconnected from the<br />

monitor.<br />

The probe position may have changed.<br />

Check the probe site.<br />

The probe is no longer in contact with the patient.<br />

Check the probe site.<br />

Press the SILENCE/RESET pushbutton to clear<br />

the message.<br />

OR<br />

Re-cycle the monitor’s power.<br />

Table 8: Error Messages on the Message Window<br />

114<br />

21-02-0298 REV. 00 12/06