<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

ERROR CLASS ERROR DESCRIPTION<br />

60 Voltage Power Supply voltage exceeds algorithm min value<br />

61 Module Reference voltage under voltage or unstable<br />

65 Module Problem saving to the Module’s EEPROM<br />

66 Module Module or Probe error detection mechanism error<br />

67 Module Problem reading or saving to Module’s EEPROM<br />

68 Module Problem reading or saving to Module’s EEPROM<br />

69 Module Problem reading or saving to Module’s EEPROM<br />

70 Probe Problem reading the probe’s EEPROM<br />

74 Module Internal Error, EEPROM not initialized<br />

75 Probe Cannot read probe’s EEPROM correctly<br />

80 Module Problem with turning probe heater off<br />

81 Module Problem with turning probe heater on<br />

82 Module Internal error<br />

83 Module Internal error<br />

84 Module Internal error<br />

85 Module Heater hardware fail safe error<br />

Table 7: SureTemp Plus® Temperature Error Codes<br />

Should any of these problems persist, contact your dealer or CAS Medical Systems, Inc.<br />

112<br />

21-02-0298 REV. 00 12/06