<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

ERROR CLASS ERROR DESCRIPTION<br />

00 Transmit buffer overflow<br />

01 Probe Probe heater energy accumulation too high<br />

02 Probe Probe a/d pulse width out of range<br />

03 Probe Adaptive probe gain too high or too low<br />

11 Ambient Ambient temperature above 104 °F<br />

12 Ambient Ambient temperature below 60.8 °F<br />

21 Module Battery voltage below error threshold of 3.0 volts<br />

31 Module RAM read/write error<br />

32 Module ROM checksum error<br />

33 Module CPU instruction error<br />

40 Module PTB resistor a/d pulse width out of range<br />

41 Module RatioCal resistor a/d pulse width out of range<br />

42 Module External ambient thermistor a/d pulse width out of range<br />

50 Module Heater circuit failure<br />

51 Module Probe heated above 112 °F<br />

52 Module Heater watchdog timeout failure<br />

60 Module PTB resistor “temperature” out of rang<br />

99 Configuration Temperature option no longer recognized<br />

Table 6: SureTemp® Temperature Error Codes<br />

Should any of these problems persist, contact your dealer or CAS Medical Systems, Inc.<br />

ERROR CLASS ERROR DESCRIPTION<br />

21 Probe Defective probe tip where heater is too close to thermistor<br />

22 Probe Excessive heater energy, heater not working<br />

24 Ambient Ambient temperature too high<br />

25 Ambient Ambient temperature too low<br />

27 Voltage Power Supply voltage exceeds algorithm max value<br />

28 Voltage Power Supply voltage below algorithm min value<br />

32 Probe Thermistor pulled away from the tip or heater broken<br />

33 Probe Probe not responsive, not characterized/calibrated<br />

34 Probe Probe not characterized/calibrated<br />

45 Prb/Mod/Pat 1 Measurement below allowable temperature values<br />

46 Prb/Mod/Pat Measurement above allowable temperature values<br />

47 Module Internal calibration resistor in module is damaged<br />

48 Module Internal calibration resistor in module is damaged<br />

49 Module Internal circuit validation resistor in module is damaged<br />

50 Module Internal circuit validation resistor in module is damaged<br />

51 Module A/D measurement timed out<br />

52 Probe Probe not characterized/calibrated<br />

59 Voltage Power Supply voltage below algorithm max value<br />

1 Prb/Mod/Pat indicates class could Probe, Module or Patient<br />

21-02-0298 REV. 00 12/06<br />