CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />
ERROR CLASS ERROR DESCRIPTION<br />
00 Transmit buffer overflow<br />
01 Probe Probe heater energy accumulation too high<br />
02 Probe Probe a/d pulse width out of range<br />
03 Probe Adaptive probe gain too high or too low<br />
11 Ambient Ambient temperature above 104 °F<br />
12 Ambient Ambient temperature below 60.8 °F<br />
21 Module Battery voltage below error threshold of 3.0 volts<br />
31 Module RAM read/write error<br />
32 Module ROM checksum error<br />
33 Module CPU instruction error<br />
40 Module PTB resistor a/d pulse width out of range<br />
41 Module RatioCal resistor a/d pulse width out of range<br />
42 Module External ambient thermistor a/d pulse width out of range<br />
50 Module Heater circuit failure<br />
51 Module Probe heated above 112 °F<br />
52 Module Heater watchdog timeout failure<br />
60 Module PTB resistor “temperature” out of rang<br />
99 Configuration Temperature option no longer recognized<br />
Table 6: SureTemp® Temperature Error Codes<br />
Should any of these problems persist, contact your dealer or CAS Medical Systems, Inc.<br />
ERROR CLASS ERROR DESCRIPTION<br />
21 Probe Defective probe tip where heater is too close to thermistor<br />
22 Probe Excessive heater energy, heater not working<br />
24 Ambient Ambient temperature too high<br />
25 Ambient Ambient temperature too low<br />
27 Voltage Power Supply voltage exceeds algorithm max value<br />
28 Voltage Power Supply voltage below algorithm min value<br />
32 Probe Thermistor pulled away from the tip or heater broken<br />
33 Probe Probe not responsive, not characterized/calibrated<br />
34 Probe Probe not characterized/calibrated<br />
45 Prb/Mod/Pat 1 Measurement below allowable temperature values<br />
46 Prb/Mod/Pat Measurement above allowable temperature values<br />
47 Module Internal calibration resistor in module is damaged<br />
48 Module Internal calibration resistor in module is damaged<br />
49 Module Internal circuit validation resistor in module is damaged<br />
50 Module Internal circuit validation resistor in module is damaged<br />
51 Module A/D measurement timed out<br />
52 Probe Probe not characterized/calibrated<br />
59 Voltage Power Supply voltage below algorithm max value<br />
1 Prb/Mod/Pat indicates class could Probe, Module or Patient<br />
21-02-0298 REV. 00 12/06<br />
111