<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />

CAUTION:<br />

This product contains a rechargeable battery that is recyclable. Under various state and local<br />

laws, it may be illegal to dispose of this battery into the municipal waste stream. Check with your<br />

local authorities for instructions on recycling options in your area.<br />

AUTO OFF FEATURE<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> incorporates an Auto “OFF” feature to preserve battery life. The Auto<br />

Off feature is enabled automatically when the user selects the “Alrm Off” option in the<br />

Configuration Menu.<br />

When the monitor is in this mode and if accidentally left powered “ON” for ten (10) minutes<br />

without any key presses or any parameters active, the monitor will beep for five (5) seconds prior<br />

to automatically shutting itself down to preserve battery life.<br />

For the Auto “OFF” feature to work properly, the following conditions must apply;<br />

• “Alrm Off” is selected in the Configuration Menu.<br />

• The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> is operating on battery power.<br />

• No NIBP, SpO 2 or Temperature readings were taken in the past ten (10) minutes.<br />

• No key presses were made in the past ten (10) minutes.<br />

• The monitor cannot be in the Automatic NIBP cycle mode.<br />

POWER FAIL<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> incorporates a Power Fail feature. Whenever the power is<br />

disconnected from the monitor and the monitor is not allowed to shut down in an orderly fashion,<br />

the monitor, when re-powered alerts the user. The message “Pwr Fail” is displayed in the<br />

Message Window and three (3) audio “beeps” followed by two (2) audio “beeps” are heard every<br />

ten (10) seconds.<br />

During this condition, all other pushbuttons are inactive except for the ON/STANDBY and<br />

SILENCE/RESET.<br />

To clear the Power Fail condition, and return to “Ready”, press the SILENCE/RESET pushbutton<br />

or recycle the monitor’s power.<br />

USER MESSAGES<br />

The <strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong> displays a variety of messages to aid the user in monitor operation.<br />

If a troubleshooting message is displayed during a measurement, follow the actions listed to<br />

correct the situation.<br />

If the monitor does not turn on, or exhibits a flashing display and failure to operate, the battery is<br />

most likely below the Dead Battery point. Connect the monitor to a power source (AC Line<br />

Power or +12 VDC) and allow it to charge for four (4) hours.<br />

108<br />

21-02-0298 REV. 00 12/06