CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
CASMED 740 Vital Signs Monitor - Implox
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />
When the message "Low Batt" appears in the Message Window, at least thirty (30) minutes of<br />
battery operation remain. The "Low Batt" message will alternate continuously with the “Ready”<br />
message to indicate that the battery should be charged as soon as possible. Also, the monitor’s<br />
front panel Battery Power Visual Indicator will change from Orange to Red and three (3) audio<br />
“beeps” are heard every twenty-five (25) seconds.<br />
WARNING:<br />
Upon the detection of a Low Battery condition and if the battery is not charged by the user, the<br />
monitor may no longer function as intended. The monitor should be plugged into a power<br />
source as soon as possible and the battery allowed to charge for four (4) hours.<br />
When the "Dead Bat" message appears, the battery is no longer able to power a measurement.<br />
The message "Dead Bat" is displayed continuously in the Message Window, the Battery Power<br />
Visual Indicator is colored Red and three (3) audio “beeps” followed by two (2) audio “beeps”<br />
once every ten (10) seconds are heard until the power is turned off.<br />
WARNING:<br />
Upon the detection of a Dead Battery condition and if the monitor is not turned off by the user,<br />
the monitor shuts down and turns “OFF” after three (3) minutes of operations.<br />
When either of these messages appears, it is necessary to recharge the battery. A depleted<br />
battery may be fully recharged in four (4) hours. The monitor can be used to obtain<br />
measurements while the battery is charging.<br />
NOTE:<br />
Using the monitor while charging may lengthen the time to restore battery charge.<br />
NOTE:<br />
During charging of the battery, the case may feel warm to the touch.<br />
106<br />
21-02-0298 REV. 00 12/06