<strong>CASMED</strong> <strong>740</strong> <strong>Monitor</strong>s<br />
MAP VALUE ENABLE / DISABLE<br />
During a blood pressure reading, the user can elect to display or not to display the MAP value.<br />
Should it need to be changed, first enter the Configuration menu. Refer to Page 98, ENTERING<br />
THE CONFIGURATION MENU.<br />
Once in the menu, use one of the NEXT programmed pushbutton keys until the Message<br />
Window displays the monitor’s current setting for the MAP value “MAP On” or “MAP Off”.<br />
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection.<br />
NOTE:<br />
When “MAP Off” is selected, MAP values are omitted from History Display and Printing as well.<br />
All alarms associated with MAP values are also disabled.<br />
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the<br />
CANCEL pushbutton to exit to the “Ready” mode.<br />
SET THE SpO 2 ALARM DELAY<br />
The delay time until an alarm is generated for %SpO 2 and Pulse Rate can be configured to be<br />
either zero (0) seconds “Delay 0S” or ten (10) seconds “Delay10S”. The default value used by<br />
the Model <strong>740</strong> <strong>Monitor</strong> is ten (10) seconds.<br />
Should it need to be changed, first enter the Configuration menu. Refer to Page 98, ENTERING<br />
THE CONFIGURATION MENU.<br />
Once in the menu, use one of the NEXT programmed pushbutton keys until the Message<br />
Window displays the monitor’s current setting for the SpO 2 Alarm Delay.<br />
Use the ARROW UP or ARROW DOWN pushbuttons to make your selection.<br />
Press one of the NEXT programmed pushbutton keys to advance onto the next menu or use the<br />
CANCEL pushbutton to exit to the “Ready” mode.<br />