Chapter 8<br />
Fiber testing (LWL) FiberTEK FDX<br />
8.4. Test Setup According to Reference Method with Three Test Cords<br />
(based on ISO/IEC 14763-3)<br />
This is the preferred measuring method, which can be used for any topology - both<br />
installation and transmission line. Irrespective of the plug face, it is also used on the<br />
measuring equipment and within the transmission line.<br />
Calibration takes place via reference line between the launch cords and tail cords used for<br />
measurement.<br />
This measurement method is applied to FiberTEK FDX.<br />
Three Test Cords<br />
Illustration 8.4. Test Setup According to Reference Method with<br />
8.5. Test setup According to Reference Method with One Test Cord<br />
(based on ISO/IEC 14763-3)<br />
This measuring method is only permitted for measurement of installation lines that are set up<br />
as Simplex lines. Furthermore, all connectors on the measuring equipment and within the line<br />
must be of the same type.<br />
Calibration is performed with the launch cord used for the measurement. Subsequently, the<br />
line to be measured and a tail cord are introduced.<br />
Illustration 8.5. Test Setup According to Reference Method with One<br />
Test Cord<br />