Chapter 5 Autotest 4. Connect Display handset (DH) and Remote handset (RH) with patch cord for Display handset (DH), as shown on the display. Remember which connection of the patch cord is connected to the Display handset (DH). Illustration 3 5.9.1. Patch Cord (DH) 5. Use soft key to continue calibration. The Remote handset (RH) indicates calibration, and a progress bar shows the status. 6. Perform the same patch cord connections to Display handset (DH) and Remote handset (RH) as during the previous test steps. Illustration 5.9. Open Patch Cord 7. On Display handset (DH) use soft key to start the last step of calibration. 8. On Remote handset (RH) use Autotest to start the last step of calibration. Illustration 4 5.9.1. Calibration Complete 9. On Display handset (DH) press Escape to return to screen. The Remote handset (RH) briefly displays that calibration was successful, and then shows ready state. 51
Chapter 5 Autotest 5.9.2. Coaxial Cable Calibration must be performed if: ‣ During the last 7 days no calibration has been performed. The user receives a message, informing him that a field calibration is required. ‣ The test adapter has been changed. ‣ A patch cord has been changed. For calibration with COAX adapters, the LanTEK ® II process, consisting of 4 steps, is changed. Since the coaxial cables are tested at low frequencies, the additional data received during calibration are largely ignored, thus calibration with LanTEK ® II Cable Tester is already completed after one step. NOTE: The LanTEK ® II Cable Tester COAX adapters are equipped with BNC connections. COAX cables applied by the user must convert from/to BNC and F (CATV) or other plug types. 1. Open "Cable Type" on start screen. 2. Use Arrow Keys to navigate to option Ethernet and confirm with Enter. 3. Use Arrow Keys to navigate to desired cable type and confirm with Enter. 4. Connect Coax adapters to Display handset (DH) and Remote handset (RH). 5. Connect the short coaxial calibration cable to the adapters of Display handset (DH) and Remote handset (RH). 6. Use Arrow Keys to navigate on ready screen to "Calibration" and press Enter. Illustration 1 5.9.2. Ready Screen 7. Use soft key to initiate calibration. Illustration 2 5.9.2. Coaxial Cable Calibration 52
- Page 1 and 2: LanTEK ® II Series Manual 0
- Page 3 and 4: Standard Warranty Policy of IDEAL I
- Page 5 and 6: Table of contents CHAPTER 1 Your La
- Page 7 and 8: Table of contents 6.11.1. Insertion
- Page 9 and 10: Table of contents 13.1. Technical S
- Page 11 and 12: Chapter 1 Your LanTEK ® II Cable C
- Page 13 and 14: CHAPTER 2 Product Description 2.1.
- Page 15 and 16: Chapter 2 Product Description 2.1.2
- Page 17 and 18: Chapter 2 Product Description 2.2.
- Page 19 and 20: 2.3. Power Management Chapter 2 Pro
- Page 21 and 22: CHAPTER 3 Basics of the Cable Test
- Page 23 and 24: Chapter 4 Preferences 1. Use Arrow
- Page 25 and 26: Chapter 4 Preferences 1. Use Arrow
- Page 27 and 28: Chapter 4 Preferences 6. Use soft k
- Page 29 and 30: CHAPTER 5 Autotest Using Autotest,
- Page 31 and 32: Chapter 5 Autotest Autotest Pref Th
- Page 33 and 34: Chapter 5 Autotest 2. Use Arrow Key
- Page 35 and 36: Chapter 5 Autotest 5.3.1. Simple Ca
- Page 37 and 38: Chapter 5 Autotest 3. Use Alphanume
- Page 39 and 40: Chapter 5 Autotest 5.4.1. Cable Nam
- Page 41 and 42: Chapter 5 Autotest AAA A / AAA A -
- Page 43 and 44: Chapter 5 Autotest Illustration 2 5
- Page 45 and 46: Chapter 5 Autotest 20. Make correct
- Page 47 and 48: Chapter 5 Autotest 5.6. DualMODE Fu
- Page 49 and 50: Chapter 5 Autotest Illustration 2 5
- Page 51: Chapter 5 Autotest 5.9. Calibration
- Page 55 and 56: Chapter 5 Autotest 3. Use soft key
- Page 57 and 58: Chapter 5 Autotest Illustration 4 5
- Page 59 and 60: CHAPTER 6 Structured Cabling Testin
- Page 61 and 62: Chapter 6 Structured cabling testin
- Page 63 and 64: Chapter 6 Structured cabling testin
- Page 65 and 66: 6.11. Insertion Loss Test (Attenuat
- Page 67 and 68: Chapter 6 Structured cabling testin
- Page 69 and 70: Chapter 6 Structured cabling testin
- Page 71 and 72: Chapter 6 Structured cabling testin
- Page 73 and 74: CHAPTER 8 Fiber Testing with FiberT
- Page 75 and 76: Chapter 8 Fiber testing (LWL) Fiber
- Page 77 and 78: Chapter 8 Fiber testing (LWL) Fiber
- Page 79 and 80: Chapter 8 Fiber testing (LWL) Fiber
- Page 81 and 82: Chapter 8 Fiber testing (LWL) Fiber
- Page 83 and 84: Chapter 8 Fiber testing (LWL) Fiber
- Page 85 and 86: Chapter 8 Fiber testing (LWL) Fiber
- Page 87 and 88: CHAPTER 9 Tone Generator 9.1. Tone
- Page 89 and 90: CHAPTER 10 IDEAL DataCENTER Softwar
- Page 91 and 92: Chapter 10 IDEAL DataCENTER Softwar
- Page 93 and 94: Chapter 10 IDEAL DataCENTER Softwar
- Page 95 and 96: Chapter 10 IDEAL DataCENTER Softwar
- Page 97 and 98: Chapter 10 IDEAL DataCENTER Softwar
- Page 99 and 100: Chapter 10 IDEAL DataCENTER Softwar
- Page 101 and 102: Chapter 10 IDEAL DataCENTER Softwar
Chapter 5<br />
Autotest<br />
4. Connect Display handset (DH) and Remote handset (RH) with patch cord for<br />
Display handset (DH), as shown on the display. Remember which connection of the<br />
patch cord is connected to the Display handset (DH).<br />
Illustration 3 5.9.1. Patch Cord (DH)<br />
5. Use soft key to continue calibration. The Remote handset (RH) indicates<br />
calibration, and a progress bar shows the status.<br />
6. Perform the same patch cord connections to Display handset (DH) and Remote<br />
handset (RH) as during the previous test steps.<br />
Illustration 5.9. Open Patch Cord<br />
7. On Display handset (DH) use soft key to start the last step of calibration.<br />
8. On Remote handset (RH) use Autotest to start the last step of calibration.<br />
Illustration 4 5.9.1. Calibration Complete<br />
9. On Display handset (DH) press Escape to return to screen. The Remote handset<br />
(RH) briefly displays that calibration was successful, and then shows ready state.<br />
51