Chapter 5<br />
Autotest<br />
5.9. Calibration<br />
A field calibration must be performed if:<br />
‣ During the last 7 days no calibration has been performed. The user receives a message,<br />
informing him that a field calibration is required.<br />
‣ The test adapter has been changed.<br />
‣ A patch cord has been changed.<br />
5.9.1. Twisted Pair Cabling<br />
1. Use Arrow Keys to navigate to "Calibration" and press Enter to open job list.<br />
Illustration 1 5.9.1. Ready screen<br />
Calibration requires one patch cord each for the Display handset (DH) and the Remote<br />
handset (RH).<br />
NOTE:<br />
It is imperative not to mix up the corresponding patch cords for Display handset (DH) and<br />
Remote handset (RH), and they must also be used for the cable test. We recommend to<br />
mark the patch cords accordingly.<br />
2. Connect Display handset (DH) and Remote handset (RH) with patch cord for<br />
Remote handset (RH), as shown on the display. Remember which connection of<br />
the patch cord is connected to the Remote handset (RH).<br />
Illustration 2 5.9.1. Patch Cord (RH)<br />
3. Use soft key to initiate calibration. The Remote handset (RH) indicates<br />
calibration, and a progress bar shows the status.<br />
50