<strong>MX600</strong>/700 Interface Specifications<br />

Measurement Connectors ODU out (Proprietary)<br />

Link (MSL) Voltage 56 V ±10 %<br />

Power<br />

45 W<br />

Power Sync. RS-422 compliant input<br />

78.125 kHz (typical)<br />

LAN signals IEEE 802.3 10-Base-T<br />

compliant<br />

Serial signals RS-422 compliant<br />

Video Interface<br />

(st<strong>and</strong>ard)<br />

Connector DVI-I (digital <strong>and</strong><br />

analog, single link)<br />

(only compatible Digital video signals single link TMDS<br />

with selected Analog video signals 0.7 Vpp@75<br />

displays) HSYNC/VSYNC TTL<br />

signals<br />

DDC signals none<br />

DDC power 5V ±5% @0-55 mA<br />

ECG Sync Output/Analog ECG Output (1/4” stereo phone jack<br />

with tip, ring, sleeve)<br />

General Connector 1/4” phone each with<br />

tip, ring, sleeve<br />

Isolation<br />

functional isolation<br />

Analog ECG<br />

Gain error