Series 1400 Pulsatile Blood Pumps Manual - Harvard Apparatus
Series 1400 Pulsatile Blood Pumps Manual - Harvard Apparatus Series 1400 Pulsatile Blood Pumps Manual - Harvard Apparatus
15 Diagrams H a r v a r d A p p a r a t u s S e r i e s 1 4 0 0 P u l s a t i l e B l o o d P u m p s U s e r ' s M a n u a l Single Phase Motor Speed Control (Models 1405 & 1407) BLACK #18 L2 BLACK BLACK DS1 F1 FUSE SLOW BLOW BROWN #18 MC1 L1 S3 S2 S1 PILOT LIGHT WHITE #18 W1 PLUG AC MALE BLUE #18 L1 A1 A2 F1 F2 S1 POWER SWITCH GREEN/YELLOW #18 BLACK #18 GREEN #18 BLACK #18 BLUE #18 GROUND LUG MAIN CHASIS RED BLUE #18 GROUND LUG FRONT PANEL Catalog No. Model No. Power M1 MOTOR VR101 GREEN 100K SPEED BLACK 55-1838 1405 115/240 VAC, 50–60 Hz (1 AMP) 55-3321 1421 115 VAC, 60 Hz (1 AMP) 55-3339 1421A 240 VAC, 50 Hz (1/2 AMP) 55-3305 1423 115 VAC, 60 Hz (1 AMP) 55-3313 1423A 240 VAC, 50 Hz (1/2 AMP) 52-9552 1407 115/240 VAC, 50–60 Hz (1 AMP)
16 Data H a r v a r d A p p a r a t u s S e r i e s 1 4 0 0 P u l s a t i l e B l o o d P u m p s U s e r ' s M a n u a l Hemolysis Test Data (Models 1421 & 1423) In these studies a reservoir of 500 to 800 cc of fresh dog blood was used, connected to the pump by 3/8" PVC tubing. Samples at room temperature were taken at 15 and 30 minute intervals for 4 to 5 hours. Samples were spun down and hemolysis measured immediately using the method of FLINK & WATSON. Since the rate of hemolysis depends on the amount of blood in the system and the flow rate, the results are reported as mg % per pass. The flow rate divided by the volume of blood in the system determines the number of passes through the pump per minute. Hemolysis ranged from 0.114 mg % to 0.27 mg % per hour pass through the various pumps with an error of ± 10%. To put these results in perspective most physiological perfusions are run with flow rates and total blood primes, so that the number of passes through the pump will range from about 1/4 to 3/4 per minute. Assuming one pass in two minutes and no physiological removal of the products of hemolysis,then hemolysis rates would range from 3.4 mg % to 8.7 mg % per hour of pump use.
- Page 1 and 2: Series 1400 Pulsatile Blood Pumps U
- Page 3 and 4: 2 Warranty and Repair Information H
- Page 5 and 6: 4 Specifications H a r v a r d A p
- Page 7 and 8: 6 Operation H a r v a r d A p p a r
- Page 9 and 10: 8 Operation: Models 1405 & 1407 H a
- Page 11 and 12: 10 Drawing and Parts List (1 cc) H
- Page 13 and 14: 12 Drawing and Parts List (30 cc) H
- Page 15: 14 Diagrams H a r v a r d A p p a r
16<br />
Data<br />
H a r v a r d A p p a r a t u s S e r i e s 1 4 0 0 P u l s a t i l e B l o o d P u m p s U s e r ' s M a n u a l<br />
Hemolysis Test Data (Models 1421 & 1423)<br />
In these studies a reservoir of 500 to 800 cc of fresh dog blood was used, connected<br />
to the pump by 3/8" PVC tubing. Samples at room temperature were taken at 15 and<br />
30 minute intervals for 4 to 5 hours. Samples were spun down and hemolysis measured<br />
immediately using the method of FLINK & WATSON. Since the rate of hemolysis<br />
depends on the amount of blood in the system and the flow rate, the results are<br />
reported as mg % per pass. The flow rate divided by the volume of blood in the system<br />
determines the number of passes through the pump per minute.<br />
Hemolysis ranged from 0.114 mg % to 0.27 mg % per hour pass through the various<br />
pumps with an error of ± 10%.<br />
To put these results in perspective most physiological perfusions are run with flow<br />
rates and total blood primes, so that the number of passes through the pump will<br />
range from about 1/4 to 3/4 per minute. Assuming one pass in two minutes and no<br />
physiological removal of the products of hemolysis,then hemolysis rates would range<br />
from 3.4 mg % to 8.7 mg % per hour of pump use.