The Broadband Noise calibration signal consists of a constant stream<br />
of white noise, which lasts for the specified number of minutes. The<br />
Period and Frequency are ignored.<br />
3.2.4 Mass control<br />
If supported by the instrument, the CMG-CD24 can initiate locking,<br />
unlocking, or centring of the mass. If the instrument does not support<br />
locking, unlocking and centring, instructing the CD24 to “centre”<br />
switches the instrument into a one-second response mode, which<br />
allows you to monitor the mass positions more easily.<br />
You can issue this instruction from the Mass Control tab:<br />
To switch into one-second response mode, click Centre.<br />
3.2.5 Data flow<br />
The CD24 operates in one of several transmission modes. These modes<br />
relate to how the unit uses its Flash memory:<br />
• as a simple data store, from which you can request data<br />
(FILING, DUAL and DUPLICATE modes);<br />
• as a buffer holding unacknowledged blocks, which are<br />
transmitted in preference to real-time data (FIFO mode); or<br />
• as a buffer holding unacknowledged blocks, which are<br />
transmitted whenever the channel is free but no real-time data<br />
blocks are ready (ADAPTIVE mode).<br />
• not at all (DIRECT mode).<br />
40 Issue C