<strong>GC</strong> CHROMATOGRAMS | PETROLEUM & PETROCHEMICAL<br />

Hydrocarbon Gases<br />

Refinery Gas Calibration Standard on Refinery Gas Packed Column Set<br />

(packed)<br />

Peaks<br />

1. C5-C6+<br />

2. Methane<br />

3. Ethane<br />

4. Ethylene<br />

5. Propane<br />

6. Acetylene<br />

7. Propylene<br />

8. Cyclopropane<br />

9. Isobutane<br />

10. Propadiene<br />

11. n-Butane<br />

12. 1-Butene<br />

13. Isobutene<br />

14. trans-2-Butene<br />

15. cis-2-Butene<br />

16. 1,3-Butadiene<br />

17. Isopentane<br />

18. n-Pentane<br />

Column Refinery gas column set (3 columns): 2a 2b 2c ,<br />

Silcosteel® tubing (cat.# 88000-875)<br />

Sample<br />

Refinery gas<br />

Injection<br />

Inj. Vol.:<br />

1,000 μL packed on-column<br />

Inj. Temp.: 150 °C<br />

Oven<br />

Oven Temp: 60 °C<br />

Carrier Gas He, constant flow<br />

Flow Rate: 30 mL/min<br />

Detector FID @ 150 °C<br />

<strong>GC</strong>_PC00374<br />

Refinery Gas on Rt®-Alumina BOND/KCl<br />

(PLOT)<br />

Peaks<br />

1. Methane<br />

2. Ethane<br />

3. Ethylene<br />

4. Propane<br />

5. Propylene<br />

6. Isobutane<br />

7. n-Butane<br />

8. Propadiene<br />

9. Acetylene<br />

10. trans-2-Butene<br />

11. 1-Butene<br />

12. Isobutylene<br />

13. cis-2-Butene<br />

14. Isopentane<br />

15. n-Pentane<br />

16. 1,3-Butadiene<br />

17. trans-2-Pentene<br />

18. 2-Methyl-2-butene<br />

19. 1-Pentene<br />

20. cis-2-Pentene<br />

21. Hexanes<br />

Column<br />

Rt®-Alumina BOND/KCl, 50 m, 0.53 mm ID, 10 μm<br />

(cat.# 19760)<br />

Sample<br />

Refinery gas standard<br />

Injection<br />

Inj. Vol.:<br />

10 μL split<br />

Liner: Taper (2 mm) (cat.# 20795)<br />

Inj. Temp.: 200 °C<br />

Split Vent<br />

Flow Rate: 80 mL/min<br />

Oven<br />

Oven Temp: 45 °C (hold 1 min) to 200 °C at 10 °C/min<br />

(hold 3.5 min)<br />

Carrier Gas H2, constant pressure (8.0 psi, 55.2 kPa)<br />

Linear Velocity: 74 cm/sec @ 45 °C<br />

Detector FID @ 200 °C<br />

Make-up Gas<br />

Type:<br />

N2<br />

Data Rate: 20 Hz<br />

Instrument HP5890 <strong>GC</strong><br />

<strong>GC</strong>_PC01088<br />

664 www.restek.com 1-800-356-1688 or 1-814-353-1300