<strong>GC</strong> CHROMATOGRAMS | PETROLEUM & PETROCHEMICAL<br />

Fuels<br />

Total Glycerin in Biodiesel on MXT®-Biodiesel TG<br />

Integra-Gap® Technology<br />

• Built-in retention gap<br />

• Eliminates connector<br />

The best solution for<br />

analyzing total glycerin<br />

in biodiesel!<br />

Retention Gap<br />

Liquid Phase<br />

<strong>GC</strong>_PC00969<br />

Column<br />

MXT®-Biodiesel TG, 14 m w/2 m Integra-Gap® (16 m total length),<br />

0.53 mm ID, 0.16 μm (cat.# 70289)<br />

Sample<br />

Biodiesel (B100), derivatized<br />

Diluent:<br />

Heptane<br />

Injection<br />

Inj. Vol.:<br />

1 μL cold on-column<br />

Temp. Program: Oven track<br />

Oven<br />

Oven Temp: 50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C<br />

at 30 °C/min (hold 5 min)<br />

Carrier Gas H2, constant flow<br />

Flow Rate: 4 mL/min<br />

Detector FID @ 380 °C<br />

Biodiesel Oil (B100) on Rtx®-Biodiesel TG<br />

A. Rtx®-Biodiesel TG<br />

with Alumaseal® connector and<br />

2 meters of 0.53 mm ID Hydroguard® tubing<br />

B. Rtx®-Biodiesel TG<br />

with universal Press-Tight® connector and<br />

2 meters of 0.53 mm ID Hydroguard® tubing<br />

<strong>GC</strong>_PC00905<br />

Column Rtx®-Biodiesel TG, 10 m, 0.32 mm ID, 0.10 μm (cat.# 10291)<br />

using Hydroguard® retention gap 2 m, 0.53 mm ID<br />

Sample<br />

Biodiesel (B100) with butanetriol<br />

Conc.:<br />

12.5 ppm (12.5 ng on-column)<br />

Injection<br />

Inj. Vol.:<br />

1.0 μL cold on-column<br />

Temp. Program: Oven track mode<br />

Oven<br />

Oven Temp: 50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C<br />

at 30 °C/min (hold 5 min)<br />

Carrier Gas H2, constant flow<br />

Flow Rate: 4 mL/min<br />

Linear Velocity: 79 cm/sec<br />

Detector FID @ 380 °C<br />