<strong>GC</strong> CHROMATOGRAMS | FORENSICS<br />

Arson Accelerants<br />

Unweathered Kerosene on Rxi®-1ms<br />

Rxi® Technology!<br />

Exceptionally inert,<br />

ultra low-bleed capillary<br />

columns.<br />

Mineral Spirits on Rxi®-1ms<br />

Rxi® Technology!<br />

Exceptionally inert, ultra low-bleed<br />

capillary columns.<br />

<strong>GC</strong>_PC01060<br />

Column Rxi®-1ms, 30 m, 0.25 mm ID, 0.25 μm (cat.# 13323)<br />

Sample Kerosene: unweathered (cat.# 31229)<br />

Diluent:<br />

Methylene chloride<br />

Conc.:<br />

5,000 μg/mL<br />

Injection<br />

Inj. Vol.: 1 μL split (split ratio 20:1)<br />

Liner: Splitless taper (4 mm) w/wool (cat.# 22405)<br />

Inj. Temp.: 250 °C<br />

Oven<br />

Oven Temp: 50 °C (hold 2 min) to 75 °C at 10 °C/min to 300 °C at 20 °C/min<br />

(hold 5 min)<br />

Carrier Gas He, constant linear velocity<br />

Linear Velocity: 28 cm/sec @ 50 °C<br />

Detector FID @ 300 °C<br />

<strong>GC</strong>_PC01066<br />

Column Rxi®-1ms, 30 m, 0.25 mm ID, 0.25 μm (cat.# 13323)<br />

Sample Mineral spirits: unweathered (cat.# 31225)<br />

Diluent:<br />

Methylene chloride<br />

Conc.:<br />

5,000 μg/mL<br />

Injection<br />

Inj. Vol.: 1.0 μL split (split ratio 20:1)<br />

Liner: Splitless taper (4 mm) w/wool (cat.# 22405)<br />

Inj. Temp.: 250 °C<br />

Oven<br />

Oven Temp: 50 °C (hold 2 min) to 75 °C at 10 °C/min to 300 °C at 20 °C/min<br />

(hold 5 min)<br />

Carrier Gas He, constant flow<br />

Linear Velocity: 28 cm/sec<br />

Detector FID @ 300 °C<br />

50% Weathered Diesel Fuel #2 on Rxi®-1ms<br />

Rxi® Technology!<br />

Exceptionally inert, ultra low-bleed<br />

capillary columns.<br />

Unweathered Diesel Fuel #2 on Rxi®-1ms<br />

Rxi® Technology!<br />

Exceptionally inert, ultra low-bleed<br />

capillary columns.<br />

<strong>GC</strong>_PC00921<br />

Column Rxi®-1ms, 30 m, 0.25 mm ID, 0.50 μm (cat.# 13338)<br />

Sample Diesel fuel #2: 50% weathered (cat.# 31235)<br />

Diluent:<br />

Methylene chloride<br />

Conc.:<br />

5,000 μg/mL<br />

Injection<br />

Inj. Vol.: 1.0 μL split (split ratio 20:1)<br />

Liner: Splitless taper (4 mm) w/wool (cat.# 22405)<br />

Inj. Temp.: 250 °C<br />

Oven<br />

Oven Temp: 50 °C (hold 2 min) to 75 °C at 10 °C/min to 300 °C at 20 °C/min<br />

(hold 5 min)<br />

Carrier Gas He, constant linear velocity<br />

Linear Velocity: 28 cm/sec @ 50 °C<br />

Detector FID @ 300 °C<br />

<strong>GC</strong>_PC00922<br />

Column Rxi®-1ms, 30 m, 0.25 mm ID, 0.50 μm (cat.# 13338)<br />

Sample Diesel fuel #2: unweathered (cat.# 31233)<br />

Diluent:<br />

Methylene chloride<br />

Conc.:<br />

5,000 μg/mL<br />

Injection<br />

Inj. Vol.: 1.0 μL split (split ratio 20:1)<br />

Liner: Splitless taper (4 mm) w/wool (cat.# 22405)<br />

Inj. Temp.: 250 °C<br />

Oven<br />

Oven Temp: 50 °C (hold 2 min) to 75 °C at 10 °C/min to 300 °C at 20 °C/min<br />

(hold 5 min)<br />

Carrier Gas He, constant linear velocity<br />

Linear Velocity: 28 cm/sec @ 50 °C<br />