30.12.2013 Views

GC Chromatograms

GC Chromatograms

GC Chromatograms

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>GC</strong> CHROMATOGRAMS | FOOD SAFETY<br />

Food Contaminants<br />

Acrylamide (Reference Standard) on Stabilwax®<br />

Acrylamide (Potato Chip Extract) on Stabilwax®<br />

Peaks<br />

1. Acrylamide<br />

Peaks<br />

1. Acrylamide<br />

1<br />

<strong>GC</strong>_FF00642<br />

0.5 1.0 1.5 2.0<br />

blank<br />

2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.5 7.0<br />

Time (min)<br />

<strong>GC</strong>_FF00643<br />

Column Stabilwax®, 15 m, 0.53 mm ID, 0.50 μm (cat.# 10637)<br />

Sample<br />

Acrylamide standard<br />

Diluent:<br />

Water<br />

Conc.:<br />

25 μg/mL<br />

Injection<br />

Inj. Vol.:<br />

1 μL splitless (hold 0.5 min)<br />

Liner: Auto SYS splitless w/wool (2 mm ID) (cat.# 20830)<br />

Inj. Temp.: 260 °C<br />

Oven<br />

Oven Temp: 100 °C (hold 0.5 min) to 200 °C at 15 °C/min<br />

Carrier Gas He, constant pressure<br />

Linear Velocity: 62 cm/sec @ 100 °C<br />

Detector FID @ 260 °C<br />

Column Stabilwax®, 15 m, 0.53 mm ID, 0.50 μm (cat.# 10637)<br />

Sample<br />

Conc.:<br />

1 g of potato chips spiked with 100 μg acrylamide<br />

Injection<br />

Inj. Vol.:<br />

1 μL splitless (hold 0.5 min)<br />

Liner: Auto SYS splitless w/wool (2 mm ID) (cat.# 20830)<br />

Inj. Temp.: 260 °C<br />

Oven<br />

Oven Temp: 100 °C (hold 0.5 min) to 200 °C at 15 °C/min<br />

Carrier Gas He, constant pressure<br />

Linear Velocity: 62 cm/sec @ 100 °C<br />

Detector FID @ 260 °C<br />

Hexachlorocyclohexanes in Sage QuEChERS Extract (Zoom of <strong>GC</strong>x<strong>GC</strong>-TOFMS Contour Plot)<br />

NEW!<br />

<strong>GC</strong>x<strong>GC</strong> with orthogonal<br />

Rxi®-5Sil MS and Rtx®-200<br />

columns is a powerful way<br />

to handle complex samples<br />

like dietary supplement<br />

extracts. Note separation<br />

of gamma-HCH from the<br />

isobaric interference above.<br />

Column Rxi®-5Sil MS 30 m, 0.25 mm ID, 0.25 μm (cat.# 13623)<br />

Rtx®-200 1.5 m, 0.18 mm ID, 0.20 μm (cat.# 45001)<br />

Sample<br />

Diluent:<br />

Toluene<br />

Injection<br />

Inj. Vol.:<br />

1 μL splitless (hold 1 min)<br />

Liner: Splitless taper (4 mm) w/wool (cat.# 22405)<br />

Inj. Temp.: 250 °C<br />

Purge Flow:<br />

40 mL/min<br />

Oven<br />

Oven Temp:<br />

Rxi®-5Sil MS: 80 °C (hold 1 min) to 310 °C at 4 °C/min (hold 1.5 min)<br />

Rtx®-200: 90 °C (hold 1 min) to 320 °C at 4 °C/min (hold 1.5 min)<br />

Carrier Gas<br />

He, constant flow<br />

Flow Rate:<br />

Modulation<br />

Modulator Temp.<br />

Offset: 25 °C<br />

Second Dimension<br />

Separation Time:<br />

1.8 mL/min<br />

4 sec<br />

<strong>GC</strong>_FF1194<br />

Hot Pulse Time: 1.2 sec<br />

Cool Time<br />

between Stages: 0.8 sec<br />

Detector<br />

TOFMS<br />

Transfer Line Temp.: 290 °C<br />

Analyzer Type: TOF<br />

Source Temp.: 225 °C<br />

Electron Energy: 70 eV<br />

Mass Defect: -20 mu/100 u<br />

Solvent Delay Time: 4 min<br />

Ionization Mode: EI<br />

Acquisition Range: 45 to 550 amu<br />

Spectral Acquisition<br />

Rate:<br />

100 spectra/sec<br />

Instrument<br />

LECO Pegasus 4D <strong>GC</strong>x<strong>GC</strong>-TOFMS<br />

Notes<br />

See application note GNAN1338 for extraction and cleanup details. A 1.5 m<br />

length of the Rtx®-200 column was trimmed from a 10 m column. Columns<br />

were connected with a universal Press-Tight® connector (cat.# 20429).<br />

628 www.restek.com 1-800-356-1688 or 1-814-353-1300

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!