<strong>GC</strong> CHROMATOGRAMS | PETROLEUM & PETROCHEMICAL<br />
Permanent Gases<br />
Permanent Gases on ShinCarbon ST<br />
(micropacked)<br />
Permanent Gases and C1+C2 Hydrocarbons<br />
on ShinCarbon ST<br />
(micropacked)<br />
2<br />
3 4<br />
5<br />
Peaks<br />
1. Hydrogen<br />
2. Oxygen<br />
3. Nitrogen<br />
4. Carbon monoxide<br />
5. Methane<br />
6. Carbon dioxide<br />
Column ShinCarbon ST, 100/120 mesh, 2 m,<br />
1 /16 in. OD, 1.0 mm ID (cat.# 19808)<br />
Sample<br />
Permanent gases mix, approx. 5 mol % each<br />
Injection<br />
Inj. Vol.:<br />
5 μL packed on-column<br />
Inj. Temp.: 100 °C<br />
Oven<br />
Oven Temp: 40 °C (hold 3 min) to 250 °C at 8 °C/min<br />
(hold 10 min)<br />
Carrier Gas He, constant flow<br />
Flow Rate: 10 mL/min<br />
Detector HID @ 200 °C<br />
1<br />
2<br />
Peaks<br />
1. Oxygen<br />
2. Nitrogen<br />
3. Carbon monoxide<br />
4. Methane<br />
5. Carbon dioxide<br />
6. Acetylene<br />
7. Ethylene<br />
8. Ethane<br />
Column ShinCarbon ST, 100/120 mesh, 2 m,<br />
1<br />
/16 in. OD, 1.0 mm ID (cat.# 19808)<br />
Sample<br />
Permanent gases-C1+C2 hydrocarbons<br />
Injection<br />
Inj. Vol.:<br />
5 μL packed on-column<br />
Inj. Temp.: 100 °C<br />
Oven<br />
Oven Temp: 40 °C (hold 3 min) to 250 °C at 8 °C/min<br />
(hold 10 min)<br />
Carrier Gas He, constant flow<br />
Flow Rate: 10 mL/min<br />
Detector HID @ 200 °C<br />
1<br />
6<br />
3<br />
4<br />
5 6<br />
7<br />
8<br />
0<br />
5<br />
10 15 20<br />
Time (min)<br />
<strong>GC</strong>_PC00666<br />
0<br />
5<br />
10 15 20<br />
Time (min)<br />
<strong>GC</strong>_PC00665<br />
Permanent Gases on Molecular Sieve 5A and Molecular Sieve 13X<br />
(packed)<br />
Peaks<br />
1. Oxygen<br />
2. Nitrogen<br />
3. Methane<br />
4. Carbon monoxide<br />
Columns Molesieve 5A 1 m, 2 mm ID (cat.# 80440-800)<br />
and Molesieve 13X 2 m, 2 mm ID (cat.# 80439-800)<br />
Sample<br />
5-10% each component balance helium<br />
Injection<br />
Inj. Vol.:<br />
10 μL direct<br />
Inj. Temp.: 150 °C<br />
Oven<br />
Oven Temp: 50 °C<br />
Carrier Gas H2, constant flow<br />
Flow Rate: 30 mL/min<br />
Detector TCD @ 200 °C<br />
Molecular Sieve 5A:<br />
Excellent resolution of oxygen, nitrogen,<br />
methane, and carbon monoxide.<br />
Molecular Sieve 13X:<br />
Rapid analysis and excellent peak symmetry<br />
for carbon monoxide.<br />
Excellent peak<br />
shape.<br />
<strong>GC</strong>_PC00166<br />