Radiation Curing Test Methods - infoHouse
Radiation Curing Test Methods - infoHouse Radiation Curing Test Methods - infoHouse
D 2979 5.3 Forge Gage-A spring device with a dial indicator that retains the maximum force reading until reset manually. The spring characteristics shall be such that between 8.9 and 22.2 N (2'and 5 lb) are required to extend it the permissible 2.5 mm (0.10 in.). The probe shall be mounted directly on the force gage. NOTE 3-0ther force-measuring devices such as strain gage load cells, or devices with different forcedeflection characteristics may be used in certain instances. Tack values obtained with these devices will differ in magnitude but will be related to standard values obtained with the specified gage. 5.4 Testing Machin&-A mechanical system for bringing the adhesive into contact with the probe, automatically controlling the dwell time during which the adhesive and probe are in contact under pressure, and subsequently pulling the adhesive away from the probe. The machine shall be capable of maintaining a constant crosshead speed of 10 f 0.1 mm/s (24 in./min), sensing contact of probe with adhesive, stopping for 1 f 0.01 s, and then reversing at the same 10 f 0.1- mm/s speed. The machine must support the probe such that its top surface is parallel to the plane of the adhesive at the time of contact to less than a 0" 10 min angle between them. NOTE 4-For some special purposes it may be desirable to measure the tack at dwell times as short as 0.1 s or as long as 100 s, depending on how quickly the adhesive bond is to be established in the end use. 6. Test Specimens 6.1 Adhesive shall be cast from solution, dispersion, or as hot melt, etc. on 38-pm (1.5-mil) polyester film to give a dried adhesive thickness, preferably between 0.04 and 0.25 mm (0.0015 and 0.01 in.), but always greater than 0.025 mm (0.001 in.) and less than 0.5 mm (0.02 in.). A minimum of two adhesive films shall be made, one from each of two separately prepared but identical solutions. The adhesive film shall be dried in a dust-free atmosphere to constant weight before testing. NOTE 5-Other films may also be used provided they do not react with the adhesive or elongate excessively during the test. Microscope slides or cover glasses may be used to give rigid support to the adhesive. 6.2 Adhesives already on some supporting material shall be examined as they exist. 7. conditioning 7.1 Testing Room-Test in a standard labo- ratory atmosphere, in accordance with Specification E 171. 8. procedure 8.1 Clean the probe with an absorbent material such as surgical gauze or tissue wet with a completely volatile solvent that is a good solvent for the adhesive under test. To be suitable, the material must be lint-free during use, absorbent, and contain no additives that are soluble in the cleaning solvents used. Wipe with another piece of solvent-wet absorbent material, and finally wipe the probe with a clean dry piece of absorbent material to remove excess solvent. Wait 20 s or a sufficiently longer time if necessary, to allow for complete evaporation of the solvent and temperature equilibration. 8.2 Place an appropriately sized sample of supported adhesive, sticky side down, on the annular ring weight. The sample need only be large enough to cover the hole in the weight without slippage during the test and should be small enough so that it does not adhere to the carrier supporting the weight. Place the weight in the carrier. NOTE 6-The proper relation of the supported adhesive on the ring weight to the probe is sketched in Fig. 1. One possible arrangement of the canier and force gage is also illustrated. 8.3 At a speed of 10 f 0.1 mm/s, bring the probe into contact with the adhesive. After a dwell time of 1.0 f 0.01 s separate the probe from the adhesive at 10 f 0.1 mm/s. 8.4 Record the tack as the maximum force in newtons required to separate the probe from the adhesive. 8.5 Make at least five determinations taken at random points on the supported adhesive. With grossly rough or nonuniform adhesive surfaces more than five determinations may be necessary to obtain the desired statistical reliability of the average tack value. 8.6 Clean the probe surface with solvent after each test. In instances where tack values neither tend to increase or decrease systematically, cleaning may be limited to each series of test on a given adhesive. Nom 7-If there is evidence of contamination by lint, large dust particles, thumb prints, etc., the test shall 'An example of a suitable machine is the Polyken Robe Tack Tester availabk from Testing Machines, Inc., Mineoh, Long Island. NY. 3 14
D 2979 be discounted. If the probe surface shows to the unaided eye presence of deposited adhesive, the test shall also be discounted. 9. Report 9.1 The report shall include the following: 9.1.1 Identification of the adhesive, 9.1.2 Solvent used for dissolving the adhesive, 9.1.3 Percent solids contained in solution, 9.1.4 Temperature and time of drying used, 9.1.5 Conditioning time for the adhesive sur- face, 9.1.6 Adhesive thickness, 9.1.7 Solvent used to clean probe, 9.1.8 Supporting backing material for the adhesive, 9.1.9 Separation rate of the probe, 9.1.10 Dwell time, 9.1.1 1 Contact pressure, 9.1.12 Temperature and relative humidity if different from standard laboratory astmosphere, and 9.1.13 Values of each of the tack readings, and the arithmetic average. Tack values shall be reported in newtons. NOTE 7-Items 9.1.2, 9.1.3, and 9.1.4 need be reported only if the adhesive is put in solution by the person making the tack test. 315
- Page 276 and 277: elative humidity of 50 f 2 percent
- Page 278 and 279: D 2095 FIG. 2 Test Specimen with At
- Page 280 and 281: 4.2 Glass Plates-Glass panels appro
- Page 282 and 283: for this purpose. * Bakers AA made
- Page 284 and 285: D 2196 having eight rotational spee
- Page 286 and 287: D 2196 speed as shown in Section 9.
- Page 288 and 289: D 2196 TABLE 1 Factors for Converti
- Page 290 and 291: this test method and any fundamenta
- Page 292 and 293: ~~ D2240 Type A method. Table 2 giv
- Page 294 and 295: em Designation: D 2293 - 69 (Reappr
- Page 296 and 297: @ D2293 1 3 3.97 mm (5132") 7 63.5
- Page 298 and 299: Lbsignatim 0 2294 - 69 (Reapproved
- Page 300 and 301: LT FIG. 1 Tc~~sion Creep Test Appar
- Page 302 and 303: 4slb Designation: D 2556 - 69 386 A
- Page 304 and 305: APPENDIX Al. EXAMPLE OF TABLE FOR V
- Page 306 and 307: D 2849 cated, references to water s
- Page 308 and 309: the galvanometer needle is constant
- Page 310 and 311: D 2849 duce a weighed quantity of t
- Page 312 and 313: D 2849 where: Ntl = Nt2 = tl = t2 =
- Page 314 and 315: D 2849 weighed to the nearest 0.1 m
- Page 316 and 317: D 2849 58. Procedure 58.1 Add 50 mL
- Page 318 and 319: D 2849 66. I .2 Calculate the equiv
- Page 320 and 321: D 2849 yethers whose viscosity fall
- Page 322 and 323: D 2849 94.3 Hvdrochloric Acid (1+17
- Page 324 and 325: ~ TABLE 3 Gardner Reference Standar
- Page 328 and 329: f A-Adhesive &Backing C--carrier W-
- Page 330 and 331: D 3111 6. Test Specimen 6.1 The tes
- Page 332 and 333: FIG. 2 Test Specimen. 1.25 2 0.01MM
- Page 334 and 335: D 3166 specified conditions of prep
- Page 336 and 337: Designation: D 3247 - 73 R84 AMERIC
- Page 338 and 339: D 3247 8.5 Place the sled with the
- Page 340 and 341: Designation: D 3248 - 73 ~ 8 4 AMER
- Page 342 and 343: D 3248 essentially constant for rep
- Page 344 and 345: lsTr, Designation: D3334 - 80 AMERI
- Page 346 and 347: 9.2 Report: 9.2. I State that the s
- Page 348 and 349: D 3334 mometer bulb 2 in. (50 mm) f
- Page 350 and 351: ~ ~ selvage than one tenth of the w
- Page 352 and 353: 0 3359 4.3 Tape-One-inch (25") wide
- Page 354 and 355: D 3359 using the illuminated magnif
- Page 356 and 357: ClSTb Designation: D 3363 - 74 (Rea
- Page 358 and 359: AMERICAN SOCIETY FOR TESTING AND MA
- Page 360 and 361: of two alternative configurations u
- Page 362 and 363: T I 0 4 190.6 m + L 4- 95.3 m + L -
- Page 364 and 365: Designation: D 3658 - 78 (Reapprove
- Page 366 and 367: D 3658 Details 1 -Base plate 2 -Upr
- Page 368 and 369: UV RADIATION Details 1 - UV specime
- Page 370 and 371: D 3807 conditions of preparation an
- Page 372 and 373: FIG. 1 Cleavage Peel Test Specimen
- Page 374 and 375: ~ wheels No. CS- 10 or CS- 17, as r
D 2979<br />
5.3 Forge Gage-A spring device with a dial<br />
indicator that retains the maximum force reading<br />
until reset manually. The spring characteristics<br />
shall be such that between 8.9 and 22.2 N (2'and<br />
5 lb) are required to extend it the permissible 2.5<br />
mm (0.10 in.). The probe shall be mounted directly<br />
on the force gage.<br />
NOTE 3-0ther force-measuring devices such as<br />
strain gage load cells, or devices with different forcedeflection<br />
characteristics may be used in certain instances.<br />
Tack values obtained with these devices will<br />
differ in magnitude but will be related to standard<br />
values obtained with the specified gage.<br />
5.4 <strong>Test</strong>ing Machin&-A mechanical system<br />
for bringing the adhesive into contact with the<br />
probe, automatically controlling the dwell time<br />
during which the adhesive and probe are in contact<br />
under pressure, and subsequently pulling the<br />
adhesive away from the probe. The machine shall<br />
be capable of maintaining a constant crosshead<br />
speed of 10 f 0.1 mm/s (24 in./min), sensing<br />
contact of probe with adhesive, stopping for 1 f<br />
0.01 s, and then reversing at the same 10 f 0.1-<br />
mm/s speed. The machine must support the<br />
probe such that its top surface is parallel to the<br />
plane of the adhesive at the time of contact to<br />
less than a 0" 10 min angle between them.<br />
NOTE 4-For some special purposes it may be desirable<br />
to measure the tack at dwell times as short as<br />
0.1 s or as long as 100 s, depending on how quickly the<br />
adhesive bond is to be established in the end use.<br />
6. <strong>Test</strong> Specimens<br />
6.1 Adhesive shall be cast from solution, dispersion,<br />
or as hot melt, etc. on 38-pm (1.5-mil)<br />
polyester film to give a dried adhesive thickness,<br />
preferably between 0.04 and 0.25 mm (0.0015<br />
and 0.01 in.), but always greater than 0.025 mm<br />
(0.001 in.) and less than 0.5 mm (0.02 in.). A<br />
minimum of two adhesive films shall be made,<br />
one from each of two separately prepared but<br />
identical solutions. The adhesive film shall be<br />
dried in a dust-free atmosphere to constant<br />
weight before testing.<br />
NOTE 5-Other films may also be used provided<br />
they do not react with the adhesive or elongate excessively<br />
during the test. Microscope slides or cover glasses<br />
may be used to give rigid support to the adhesive.<br />
6.2 Adhesives already on some supporting<br />
material shall be examined as they exist.<br />
7. conditioning<br />
7.1 <strong>Test</strong>ing Room-<strong>Test</strong> in a standard labo-<br />
ratory atmosphere, in accordance with Specification<br />
E 171.<br />
8. procedure<br />
8.1 Clean the probe with an absorbent material<br />
such as surgical gauze or tissue wet with a<br />
completely volatile solvent that is a good solvent<br />
for the adhesive under test. To be suitable, the<br />
material must be lint-free during use, absorbent,<br />
and contain no additives that are soluble in the<br />
cleaning solvents used. Wipe with another piece<br />
of solvent-wet absorbent material, and finally<br />
wipe the probe with a clean dry piece of absorbent<br />
material to remove excess solvent. Wait 20 s or<br />
a sufficiently longer time if necessary, to allow<br />
for complete evaporation of the solvent and temperature<br />
equilibration.<br />
8.2 Place an appropriately sized sample of<br />
supported adhesive, sticky side down, on the<br />
annular ring weight. The sample need only be<br />
large enough to cover the hole in the weight<br />
without slippage during the test and should be<br />
small enough so that it does not adhere to the<br />
carrier supporting the weight. Place the weight in<br />
the carrier.<br />
NOTE 6-The proper relation of the supported adhesive<br />
on the ring weight to the probe is sketched in<br />
Fig. 1. One possible arrangement of the canier and<br />
force gage is also illustrated.<br />
8.3 At a speed of 10 f 0.1 mm/s, bring the<br />
probe into contact with the adhesive. After a<br />
dwell time of 1.0 f 0.01 s separate the probe<br />
from the adhesive at 10 f 0.1 mm/s.<br />
8.4 Record the tack as the maximum force in<br />
newtons required to separate the probe from the<br />
adhesive.<br />
8.5 Make at least five determinations taken at<br />
random points on the supported adhesive. With<br />
grossly rough or nonuniform adhesive surfaces<br />
more than five determinations may be necessary<br />
to obtain the desired statistical reliability of the<br />
average tack value.<br />
8.6 Clean the probe surface with solvent after<br />
each test. In instances where tack values neither<br />
tend to increase or decrease systematically, cleaning<br />
may be limited to each series of test on a<br />
given adhesive.<br />
Nom 7-If there is evidence of contamination by<br />
lint, large dust particles, thumb prints, etc., the test shall<br />
'An example of a suitable machine is the Polyken Robe<br />
Tack <strong>Test</strong>er availabk from <strong>Test</strong>ing Machines, Inc., Mineoh,<br />
Long Island. NY.<br />
3 14