Chapter 3 Structured Cable Field Calibration and Testing GG45 – TROUBLESHOOTING A field calibration should be performed prior to implementing a test of the GG45 cabling system. This process will ensure (1) synchronizing of the units, (2) qualifying (testing) of the patchcords and (3) gathering of loss data regarding the patchcords and mated connections. The equipment required for a GG45 calibration procedure is: ‣ Display Handset ‣ Remote Handset ‣ GG45 Calibration Adapters (Two adapters joined by a short segment of Category 7 cable) ‣ GG45 Calibration Load Terminator (100Ω Jack) ‣ GG45 Permanent Link Adapters (A set of two adapters, each with a patchcord (approximately 2 meters) soldered to the adapter at one end and a GG45 Category 7 plug at the opposite end) TESTING A GG45 CABLING SYSTEM Figure 3-1: Typical Configuration for GG45 Cable Testing Typical configuration for testing has the DH unit and RH unit connected to the GG45 Permanent Link Adapters. The cable under test is connected at the Category 7 plug end of both the DH and RH ends. 3-4

Chapter 3 Structured Cable Field Calibration and Testing FIELD CALIBRATION - PATCHCORD TESTING This Field Calibration is a 4-step process. Steps 1 and 2 are performed with the hybrid (RJ45 Plug to RJ45-TIA-Reference-Jack) patchcords connected to the Handsets using the kit’s mini patchcord. Steps 3 and 4 are performed with open-ended hybrid patchcords (Only one end connected to the Handsets). To calibrate the tester, perform the following: Step 1 ‣ Connect the Category 5e/6 – RJ45 adapters to the Display Handset (DH) and Remote Handset (RH). Power both units on. ‣ Connect the RJ45 plug of the first hybrid patchcord to the Remote Handset (RH). Using the RJ45 mini patchcord attach the DH RJ45-TIA Reference Jack end of the hybrid patchcord to one end of the mini patchcord and then connect the other end of the mini patchcord to the Display Handset (DH). Note: Patchcords can vary in length. The most common length used is the 2-meter length patchcords. ‣ From the DH Ready Screen, select Field Calibration. The Calibration screen appears. ‣ From the RH, press AUTOTEST to perform the first step of the calibration process. At the completion of step 1, disconnect the mini patchcord and hybrid patchcord from the RH and DH. Step 2 ‣ Connect the RJ45 plug of the second hybrid patchcord to the Display Handset (DH). Using the RJ45 mini patchcord attach the DH RJ45-TIA Reference Jack end of the hybrid patchcord to one end of the mini patchcord and then connect the other end of the mini patchcord to the Remote Handset (RH). ‣ From the DH Field Calibration screen, select START or press AUTOTEST from the RH to perform the second step of the calibration process. ‣ At the completion of step 2, disconnect the mini patchcord. 3-5

Chapter 3<br />

Structured Cable Field Calibration and Testing<br />

FIELD CALIBRATION - PATCHCORD TESTING<br />

This Field Calibration is a 4-step process. Steps 1 and 2 are performed with the hybrid<br />

(RJ45 Plug to RJ45-TIA-Reference-Jack) patchcords connected to the Handsets using<br />

the kit’s mini patchcord. Steps 3 and 4 are performed with open-ended hybrid<br />

patchcords (Only one end connected to the Handsets).<br />

To calibrate the tester, perform the following:<br />

Step 1<br />

‣ Connect the Category 5e/6 – RJ45 adapters to the Display Handset (DH) and<br />

Remote Handset (RH). Power both units on.<br />

‣ Connect the RJ45 plug of the first hybrid patchcord to the Remote Handset (RH).<br />

Using the RJ45 mini patchcord attach the DH RJ45-TIA Reference Jack end of the<br />

hybrid patchcord to one end of the mini patchcord and then connect the other<br />

end of the mini patchcord to the Display Handset (DH).<br />

Note: Patchcords can vary in length. The most common length used is the 2-meter<br />

length patchcords.<br />

‣ From the DH Ready Screen, select Field Calibration. The Calibration screen<br />

appears.<br />

‣ From the RH, press AUTOTEST to perform the first step of the calibration<br />

process. At the completion of step 1, disconnect the mini patchcord and hybrid<br />

patchcord from the RH and DH.<br />

Step 2<br />

‣ Connect the RJ45 plug of the second hybrid patchcord to the Display Handset<br />

(DH). Using the RJ45 mini patchcord attach the DH RJ45-TIA Reference Jack<br />

end of the hybrid patchcord to one end of the mini patchcord and then connect<br />

the other end of the mini patchcord to the Remote Handset (RH).<br />

‣ From the DH Field Calibration screen, select START or press AUTOTEST from<br />

the RH to perform the second step of the calibration process.<br />

‣ At the completion of step 2, disconnect the mini patchcord.<br />