S<br />

Innehåll<br />

sid<br />

1 Inledning<br />

1.1 Allmänt 2<br />

1.2 Tekniska data 2<br />

1.3 Termer och defi nitioner 2<br />

1.4 Säkerhets<strong>för</strong>eskrifter 2<br />

1.5 Garanti 2<br />

2 Tillvägagångssätt<br />

2.1 Spontanandning 3<br />

2.2 Bruksanvisning 3<br />

3 Service<br />

3.1 Rutinunderhåll 3<br />

3.2 Underhåll 3<br />

3.3 Rengöring av demandventilen 4<br />

3.4 Tillvägagångssätt vid rengöringen 4<br />

4 Reservdelar och tillbehör till Mediline ventilator<br />

med demandventil 4<br />

5 Återvinning 4<br />

1. Inledning<br />

1.1 Beskrivning<br />

• Mediline demandventil <strong>för</strong> <strong>oxygen</strong> är ett säkert och effektivt hjälpmedel <strong>för</strong> utbildad<br />

personal <strong>för</strong> att till<strong>för</strong>a <strong>oxygen</strong> vid inandning.<br />

• Mediline demandventil <strong>för</strong> <strong>oxygen</strong> är lätt, bärbar och mycket hållbar. Den har utformats<br />

i enlighet med de krav som ställs vid användning i prehospital vård.<br />

1.2 Tekniska data<br />

Manuellt fl öde: 0-160 liter/min ± 10%<br />

Tryckkänslighet <strong>för</strong> ventilöppning: 100 liter/min @ < -5 cm H 2<br />

O ±10%<br />

Minimalt ingångstryck: 3 bar<br />

Maximalt utgångstryck: 5 bar<br />

Tryckavlastning :<br />

45 cm H 2<br />

O<br />

Övertryckslarm:<br />

konstant ljudsignal<br />

Drifts-/<strong>för</strong>varingstemperatur: -37 °C till +50 °C<br />

Ingångsanslutning: 9/16”<br />

Patientanslutning: 15/22 mm<br />

Vikt: 227 g<br />

1.3 Termer och definitioner<br />

• <strong>Demandventil</strong>: En ventil som levererar gas till patienten med en fl ödes- hastighet<br />

motsvarande den som indikeras av patientens inandning.<br />

• Utandningsfas: Tiden från slutet på inandningsfl ödet till slutet på utandnings- fl ödet.<br />

• Utandningstid: Den tid utandningen varar.<br />

• <strong>Oxygen</strong>koncentration: <strong>Oxygen</strong>innehållet i de gaser som inandats uttryckt i procent.<br />

• Minutvolym (Vm): Den totala gasvolym som levereras till patienten under en minut.<br />

• Tryckavlastningsventil: Ventil som begränsar lungans maximala utvidgnings- tryck<br />

genom att tömma ut överfl ödsgas till atmosfär.<br />

• Andningsvolym (Vt): Den gasvolym som levereras till patienten under varje inandningsfas.<br />

1.4 Säkerhets<strong>för</strong>eskrifter<br />

Mediline demandventil <strong>för</strong> <strong>oxygen</strong> är avsedd att användas av lämpligt utbildad personal.<br />

Följande <strong>för</strong>siktighetsåtgärder bör alltid iakttas:<br />

1. Rök inte när apparaten är i bruk. Vistas inte heller nära öppen låga vare sig då apparaten<br />

används eller vid fl askbyte.<br />

2. Stäng alltid gasfl askan när apparaten inte används.<br />

8 Ensure that the cylinder is turned OFF after each <strong>use</strong> and remaining gas released<br />

from the system by slightly loosening the regulator/demand valve connection until<br />

gas fl ow ceases. Always retighten the connection after bleeding the surplus pressure.<br />

3 Servicing<br />

3.1 Routine maintenance<br />

NOTE:<br />

The Mediline <strong>oxygen</strong> demand valve is designed to provide years <strong>of</strong> reliable service in<br />

all emergency situations. Failure to follow the maintenance and inspection routines<br />

properly could result in incorrect operation <strong>of</strong> the demand valve.<br />

To ensure proper operation <strong>of</strong> the demand valve regular inspection and checking <strong>of</strong><br />

the demand valve and accessories for correct function should be undertaken by a<br />

responsible member <strong>of</strong> staff on a regular basis.<br />

This check is to ensure that all <strong>of</strong> the accessories and demand valve components are<br />

present, the gas cylinder is full and that the demand valve is in working order.<br />

3.2 Maintenance<br />

Regulator working pressure should be checked at least every six months and more<br />

frequently in high <strong>use</strong> applications. Units with test pressures outside <strong>of</strong> the ranges<br />

listed in the product specifi cations should not be <strong>use</strong>d. The Mediline products are not<br />

designed for fi eld disassembly or service outside that indicated in this manual. Any<br />

malfunctioning units should be returned to the manufacturer. Unauthorized repairs will<br />

nullify the product warranty.<br />

3.3 Cleaning the demand valve<br />

Routine cleaning <strong>of</strong> the Mediline <strong>oxygen</strong> demand valve should be undertaken to<br />

maintain the equipment in a clean condition. Face masks, patient valve and diaphragm<br />

can all be disinfected using a legally marketed commercially available<br />

disinfectant, steam sterilization or ethylene oxide gas. All other components should be<br />

cleaned with a mild soap solution.<br />

NOTE: Do not attempt to clean and sterilize any components that are designated as<br />

disposable, as immersion <strong>of</strong> these items into a sterilizing solution can ca<strong>use</strong> degeneration<br />

<strong>of</strong> the materials.<br />

3.4 Cleaning procedure<br />

The <strong>oxygen</strong> demand valve must be thoroughly cleaned after each patient <strong>use</strong>.<br />

1 Ensure that the cylinder is turned <strong>of</strong>f (completely clockwise) and any<br />

remaining gas bled out <strong>of</strong> the regulator.<br />

2 Remove the face mask from the <strong>oxygen</strong> demand valve and then disconnect<br />

the demand valve from the regulator.<br />

3 Unscrew patient valve from the demand valve body being careful to ensure<br />

that the diaphragm is retained.<br />

4 Wash all components thoroughly in a mild soap solution and disinfect as<br />

required.<br />

5 The handpiece can be wiped over with a s<strong>of</strong>t cloth and mild soap solution.<br />

6 Dry all components thoroughly.<br />

7 Re-assemble and check performance.<br />

4 The Mediline demand valve replacement parts<br />

325 111 852 Patient valve diaphragm<br />

325 111 884 Hose nut, nipple with O-ring<br />

325 111 886 O-ring for hose nipple<br />

325 111 855 Universal face mask<br />

5 Recycling<br />

Metal parts shall be recycled by scrapping, plastic parts and rubber details can be<br />

2 7